Concerns have been raised about the degradation of hydroxyapatite (HA) coatings on dental implants over time. It has been speculated that the resorption of HA could produce a space between implant and bone, resulting in mechanical instability and implant failure. A screw-shaped HA-coated implant was retrieved after the abutment had fractured after a loading period of 15 years, and was evaluated histologically. At the interface with the implant, compact mature bone, with small marrow spaces, was present. Higher magnification showed that only a few osteoblasts were present. No multinucleated giant cells were present. Bone remodeling was evidenced by the presence of many secondary osteons close to the implant surface and many reversal lines. No fibrous or granulation tissues were present at the interface. The HA coating had been completely resorbed, and just 5.1% +/- 2.3% of the implant surface was covered by the HA coating. The bone-implant contact percentage was 77.6% +/- 5.1%. The almost complete resorption of the HA coating did not appear to have interfered with osseointegration. The implant was osseointegrated and was still providing adequate function; thus the HA coating resorption probably had little clinical significance.
随着时间的推移,人们对牙种植体上羟基磷灰石(HA)涂层的降解提出了担忧。据推测,HA的吸收可能会在种植体与骨之间产生间隙,导致机械不稳定和种植体失败。一枚螺旋形HA涂层种植体在15年的加载期后基台断裂后被取出,并进行了组织学评估。在与种植体的界面处,存在致密的成熟骨,骨髓腔较小。更高倍数观察显示仅存在少数成骨细胞。未发现多核巨细胞。靠近种植体表面有许多继发性骨单位和许多反转线,证明了骨重塑的存在。界面处未发现纤维组织或肉芽组织。HA涂层已完全吸收,种植体表面仅有5.1%±2.3%被HA涂层覆盖。骨-种植体接触百分比为77.6%±5.1%。HA涂层几乎完全吸收似乎并未干扰骨整合。种植体实现了骨整合且仍在提供足够的功能;因此,HA涂层吸收可能几乎没有临床意义。