Neuhann Irmingard M, Neuhann Thomas F, Szurman Peter, Koerner Sabine, Rohrbach Jens Martin, Bartz-Schmidt Karl-Ulrich
Tuebingen University Eye Hospital, Tuebingen, Germany.
J Cataract Refract Surg. 2009 Mar;35(3):593-7. doi: 10.1016/j.jcrs.2008.08.048.
We present 3 cases of opacification in the Biocomfold 92S intraocular lens (IOL), with a documented increase in the opacification over time in 1 case. Histopathological analysis revealed the opacification was caused by calcification in the IOL's optic material. The pattern of calcium deposition, however, was different in each IOL. The causative mechanisms for this complication are unclear. Further research is warranted.