三种自动化第四代HIV抗原/抗体联合检测方法在大规模献血者和临床样本筛查中的性能

Performance of three automated fourth-generation combined HIV antigen/antibody assays in large-scale screening of blood donors and clinical samples.

作者信息

Malm K, von Sydow M, Andersson S

机构信息

Department of Clinical Microbiology, Orebro University Hospital, Orebro, Sweden.

出版信息

Transfus Med. 2009 Apr;19(2):78-88. doi: 10.1111/j.1365-3148.2009.00907.x. Epub 2009 Mar 23.

Abstract

Since the introduction in the mid-1980s, HIV testing has gradually improved both in terms of sensitivity and specificity. The so-called fourth generation of tests, combined HIV antigen/antibody assays, has now been introduced. This study compares three automated combined assays with older third-generation antibody assays in large-scale screening. Serum samples from routine screening of blood and plasma donors and clinical samples were investigated for specificity evaluation. Three fourth-generation combination assays from one manufacturer were compared with three older third-generation antibody assays from the same manufacturer. More than 40 000 samples per assay were included. For sensitivity, selected panels of confirmed HIV-1- and HIV-2-positive samples as well as seroconversion samples (HIV-1) from commercial panels and also from patients who appeared during the evaluation were used. The specificities of the fourth-generation tests were 99.91% (AxSYM), 99.95% (ARCHITECT) and 99.97% (PRISM) after repeated testing. Some specificity variation between reagent batches was observed. All HIV-1-positive samples were reactive by the three fourth-generation systems. HIV-1 seroconversion samples and panels were reactive earlier than by antibody-only tests. As for HIV-2 samples, AxSYM failed to detect one (n = 40), whereas PRISM and ARCHITECT detected all (n = 16 for PRISM and n = 52 for ARCHITECT). The new HIV antigen/antibody combination assay systems were found to have high sensitivity and specificity. The instruments provided a rational and easy way of testing at large scale.

摘要

自20世纪80年代中期引入以来,HIV检测在灵敏度和特异性方面都逐渐得到了改进。所谓的第四代检测,即HIV抗原/抗体联合检测,现已问世。本研究在大规模筛查中比较了三种自动化联合检测与较旧的第三代抗体检测。对来自血液和血浆捐献者常规筛查的血清样本以及临床样本进行了特异性评估。将一家制造商的三种第四代联合检测与同一家制造商的三种较旧的第三代抗体检测进行了比较。每种检测纳入了超过40000个样本。对于灵敏度,使用了选定的经确认的HIV-1和HIV-2阳性样本组以及来自商业样本组和评估期间出现的患者的血清转化样本(HIV-1)。重复检测后,第四代检测的特异性分别为99.91%(AxSYM)、99.95%(ARCHITECT)和99.97%(PRISM)。观察到不同试剂批次之间存在一定的特异性差异。所有HIV-1阳性样本在三种第四代检测系统中均呈反应性。HIV-1血清转化样本组比仅抗体检测更早出现反应性。对于HIV-2样本,AxSYM未能检测出1个样本(共40个样本),而PRISM和ARCHITECT检测出了所有样本(PRISM为16个样本,ARCHITECT为52个样本)。新型HIV抗原/抗体联合检测系统具有高灵敏度和特异性。这些仪器提供了一种大规模检测的合理且简便的方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验