Cernat Ramona, Dobre George M, Bradu Adrian, Podoleanu Adrian Gh
Applied Optics Group, University of Kent, Canterbury, CT2 7NH, UK.
J Opt Soc Am A Opt Image Sci Vis. 2009 Apr;26(4):723-31. doi: 10.1364/josaa.26.000723.
Key noise parameters in optical coherence tomography (OCT) systems employing splitters with a nonflat spectral response are evaluated using a supercontinuum fiber laser source with a spectrum of 450 nm-1700 nm and a time domain OCT architecture based on 1300 nm fiber splitters. The spectral behavior of the splitter leading to balanced detection is measured over a range of 300 nm. Because of spectrally different signals at the balanced detector input a residual excess photon noise term results. A rigorous treatment of this noise term [Appl. Opt.43, 4802 (2004)] introduced two new quantities that take into account the spectral properties of the coupler. In this report, we have evaluated these two noise bandwidth quantities and comparatively assessed the noise behavior predicted by the classical theory with the theory based on the two new noise bandwidths. We show that under certain operating parameters, the additional excess photon noise is twice that predicted for a coupler with a flat spectral response.
使用光谱范围为450纳米至1700纳米的超连续光纤激光源以及基于1300纳米光纤分束器的时域光学相干断层扫描(OCT)架构,对采用具有非平坦光谱响应的分束器的光学相干断层扫描(OCT)系统中的关键噪声参数进行了评估。在300纳米的范围内测量了导致平衡检测的分束器的光谱行为。由于平衡探测器输入端的信号在光谱上存在差异,会产生一个残余的过量光子噪声项。对这个噪声项的严格处理[《应用光学》43, 4802 (2004)]引入了两个新的量,它们考虑了耦合器的光谱特性。在本报告中,我们评估了这两个噪声带宽量,并将经典理论预测的噪声行为与基于这两个新噪声带宽的理论进行了比较评估。我们表明,在某些操作参数下,额外的过量光子噪声是具有平坦光谱响应的耦合器预测值的两倍。