CT图像上图像噪声和条纹伪影的定量评估:z轴自动管电流调制技术与固定管电流技术的比较。

Quantitative assessment of image noise and streak artifact on CT image: comparison of z-axis automatic tube current modulation technique with fixed tube current technique.

作者信息

Imai Kuniharu, Ikeda Mitsuru, Enchi Yukihiro, Niimi Takanaga

机构信息

Department of Radiological Technology, Nagoya University School of Health Sciences, Higashi-ku, Nagoya, Japan.

出版信息

Comput Med Imaging Graph. 2009 Jul;33(5):353-8. doi: 10.1016/j.compmedimag.2009.02.003. Epub 2009 Apr 1.

Abstract

The purpose of our study is to quantitatively assess the effects of z-axis automatic tube current modulation technique on image noise and streak artifact, by comparing with fixed tube current technique. Standard deviation of CT-values was employed as a physical index for evaluating image noise, and streak artifact was quantitatively evaluated using our devised Gumbel evaluation method. z-Axis automatic tube current modulation technique will improve image noise and streak artifact, compared with fixed tube current technique, and will make it possible to significantly reduce radiation doses at lung levels while maintaining the same image quality as fixed tube current technique.

摘要

我们研究的目的是通过与固定管电流技术进行比较,定量评估z轴自动管电流调制技术对图像噪声和条纹伪影的影响。采用CT值的标准差作为评估图像噪声的物理指标,并使用我们设计的耿贝尔评估方法对条纹伪影进行定量评估。与固定管电流技术相比,z轴自动管电流调制技术将改善图像噪声和条纹伪影,并有可能在保持与固定管电流技术相同图像质量的同时,显著降低肺部水平的辐射剂量。