Desai Pranav, Himel Van
Department of Endodontics and Operative Dentistry, University of Tennessee Health Science Center, College of Dentistry, Memphis, Tennessee 38163, USA.
J Endod. 2009 Apr;35(4):545-9. doi: 10.1016/j.joen.2009.01.011.
The objective of this project was to evaluate the safety of various intracanal irrigation systems by measuring the apical extrusion of irrigant. Twenty-two single canal, extracted mature teeth were instrumented and secured through the lid of a scintillation vial to collect apically extruded irrigant. A precision syringe pump delivered controlled amounts of irrigant at a constant flow. The irrigation systems used were EndoVac Micro and Macro Cannula, EndoActivator, manual irrigation with Max-I-Probe needle, Ultrasonic Needle Irrigation, and Rinsendo. Results were analyzed by using one-way analysis of variance with Scheffé test (P < .05). The EndoVac Micro and Macro cannulae groups did not extrude irrigant, and there was no statistically significant difference between these 2 groups and the EndoActivator group. Within the groups that produced extrusion, EndoActivator extruded statistically significantly less irrigant than Manual, Ultrasonic, and Reinsendo groups. There was no statistically significant difference among Manual, Ultrasonic, and Rinsendo groups. This study showed that the EndoVac did not extrude irrigant after deep intracanal delivery and suctioning the irrigant from the chamber to full working length. EndoActivator had a minimal, although statistically insignificant, amount of irrigant extruded out of the apex when delivering irrigant into the pulp chamber and placing the tip into the canal and initiating the sonic energy of the EndoActivator. Manual, Ultrasonic, and Rinsendo groups had significantly greater amount of extrusion compared with EndoVac and EndoActivator.
本项目的目的是通过测量冲洗液的根尖挤出量来评估各种根管冲洗系统的安全性。选取22颗单根管的成熟离体牙进行预备,并通过闪烁瓶的瓶盖固定,以收集根尖挤出的冲洗液。使用精密注射泵以恒定流速输送定量的冲洗液。所使用的冲洗系统有EndoVac微型和大型套管、EndoActivator、使用Max-I-Probe针头进行手动冲洗、超声针头冲洗以及Rinsendo。采用单因素方差分析和Scheffé检验(P < 0.05)对结果进行分析。EndoVac微型和大型套管组未出现冲洗液挤出,这两组与EndoActivator组之间在统计学上无显著差异。在有冲洗液挤出的组中,EndoActivator挤出的冲洗液在统计学上显著少于手动冲洗、超声冲洗和Rinsendo组。手动冲洗、超声冲洗和Rinsendo组之间在统计学上无显著差异。本研究表明,EndoVac在将冲洗液深入输送至根管并从髓腔抽吸至工作长度后不会挤出冲洗液。当将冲洗液注入髓腔并将尖端放入根管并启动EndoActivator的声波能量时,EndoActivator有极少量冲洗液从根尖挤出,尽管在统计学上不显著。与EndoVac和EndoActivator相比,手动冲洗、超声冲洗和Rinsendo组的挤出量显著更大。