Suppr超能文献

在模拟未插管呼吸骤停患者的实验台模型中,针对气道阻塞的通气策略。

Ventilation strategies in the obstructed airway in a bench model simulating a nonintubated respiratory arrest patient.

作者信息

Herff Holger, Paal Peter, von Goedecke Achim, Mitterlechner Thomas, Schmittinger Christian A, Wenzel Volker

机构信息

Department of Anesthesiology and Critical Care Medicine, Innsbruck Medical University, Innsbruck, Austria.

出版信息

Anesth Analg. 2009 May;108(5):1585-8. doi: 10.1213/ane.0b013e31819f32a6.

Abstract

BACKGROUND

The Smart Bag MO(R) is an adult flow-limited bag-valve device designed to reduce the risk of stomach inflation in an unprotected airway. Its properties in severe airway obstruction are as yet unknown.

METHODS

In a bench model, we evaluated respiratory mechanics and delivered tidal volumes although ventilating at airway resistances of 4, 10, and 20 cm H(2)O . L(-1) . s(-1) once with a flow-limited bag-valve device and once with a standard bag-valve device to simulate a respiratory arrest patient with an unprotected airway.

RESULTS

Inspiratory times were always longer with the flow-limited bag-valve device than with the standard bag-valve device. Lung tidal volume in the simulated unobstructed airway was 750 +/- 70 mL using the flow-limited bag-valve device versus 780 +/- 30 mL using the standard bag-valve device (n.s.); in the simulated medium obstructed airway it was 800 +/- 70 versus 850 +/- 20 mL (n.s.), and in the simulated severely obstructed airway it was 210 +/- 20 versus 170 +/- 10 mL (P < 0.01). Peak airway pressure in the simulated unobstructed airway was 15 +/- 2 cm H(2)O using the flow-limited bag-valve device versus 22 +/- 4 cm H(2)O using the standard bag-valve device (P < 0.01); in the simulated medium obstructed airway it was 22 +/- 1 versus 39 +/- 7 cm H(2)O (P < 0.01), and in the simulated severely obstructed airway it was 26 +/- 1 versus 61 +/- 3 cm H(2)O (P < 0.01). Stomach inflation in the simulated unobstructed airway was 0 mL/min using both bag-valve devices; in the simulated medium obstructed airway it was 0 mL/min for the flow-limited bag-valve device versus 200 +/- 20 mL/min for the standard bag-valve device (P < 0.01), and in the simulated severely obstructed airway it was 0 versus 1240 +/- 50 mL/min (P < 0.01).

CONCLUSION

In a simulated severely obstructed unprotected airway, the use of a flow-limited bag-valve device resulted in longer inspiratory times, higher tidal volumes, lower inspiratory pressures, and no stomach inflation compared with a standard bag-valve device.

摘要

背景

Smart Bag MO(R)是一种成人流量限制型袋阀装置,旨在降低未受保护气道发生胃胀气的风险。其在严重气道阻塞情况下的性能尚不清楚。

方法

在一个实验台模型中,我们评估了呼吸力学并在气道阻力为4、10和20 cm H₂O·L⁻¹·s⁻¹时通过流量限制型袋阀装置和标准袋阀装置各进行一次通气,以模拟气道未受保护的呼吸骤停患者,从而输送潮气量。

结果

流量限制型袋阀装置的吸气时间总是比标准袋阀装置长。在模拟的通畅气道中,使用流量限制型袋阀装置时肺潮气量为750±70 mL,而使用标准袋阀装置时为780±30 mL(无显著差异);在模拟的中度阻塞气道中,分别为800±70 mL和850±20 mL(无显著差异),在模拟的严重阻塞气道中,分别为210±20 mL和170±10 mL(P<0.01)。在模拟的通畅气道中,使用流量限制型袋阀装置时气道峰值压力为15±2 cm H₂O,而使用标准袋阀装置时为22±4 cm H₂O(P<0.01);在模拟的中度阻塞气道中,分别为22±1 cm H₂O和39±7 cm H₂O(P<0.01),在模拟的严重阻塞气道中,分别为26±1 cm H₂O和61±3 cm H₂O(P<0.01)。在模拟的通畅气道中,两种袋阀装置的胃胀气均为0 mL/min;在模拟的中度阻塞气道中,流量限制型袋阀装置的胃胀气为0 mL/min,而标准袋阀装置为200±20 mL/min(P<0.01),在模拟的严重阻塞气道中,分别为0 mL/min和1240±50 mL/min(P<0.01)。

结论

在模拟的严重阻塞且未受保护的气道中,与标准袋阀装置相比,使用流量限制型袋阀装置可导致吸气时间更长、潮气量更高、吸气压力更低且无胃胀气。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验