Klase Daniel, Bischof Arpad, Haendler Goetz, Spuck Sebastian, Rasche Dirk, Tronnier Volker
Department of Neurosurgery, Neurorothenbaum, Hamburg, Germany.
Acta Neurochir (Wien). 2009 Jun;151(6):663-7; discussion 667. doi: 10.1007/s00701-009-0299-0. Epub 2009 Apr 17.
The authors investigated the possibility of improving positioning of stimulation leads in patients with chronic neuropathic peripheral nerve pain and good pain relief from implantation of a peripheral nerve stimulator (PNS).
This pilot study includes four patients suffering from Chronic Regional Pain Syndrome type II (CRPS II) or neuropathic mononeuropathy treated with PNS therapy. The affected extremities and corresponding implantation sites were examined using computer tomographic scans (CT), additional CT angiography (CTA), reconstruction techniques and postprocessing procedures.
It was possible to prove a close relation between the implanted device and the neurovascular bundle in each of these cases. Thus, indirect lead position control was obtained.
Computer tomographic techniques represent a reliable method for the position control of implanted peripheral nerve electrodes. Hence, this procedure should surpass general radiographies in detecting lead displacements.
作者研究了改善慢性神经性周围神经痛患者刺激电极定位的可能性,这些患者通过植入外周神经刺激器(PNS)获得了良好的疼痛缓解效果。
这项前瞻性研究纳入了4例接受PNS治疗的II型慢性区域疼痛综合征(CRPS II)或神经性单神经病患者。使用计算机断层扫描(CT)、额外的CT血管造影(CTA)、重建技术和后处理程序对患侧肢体及相应植入部位进行检查。
在每例病例中均证实植入装置与神经血管束之间存在密切关系。由此获得了间接的电极位置控制。
计算机断层扫描技术是植入外周神经电极位置控制的可靠方法。因此,该方法在检测电极移位方面应优于普通X线摄影。