Feasibility and reproducibility of left ventricular rotation parameters measured by speckle tracking echocardiography.

作者信息

van Dalen Bas M, Soliman Osama I I, Vletter Wim B, Kauer Floris, van der Zwaan Heleen B, ten Cate Folkert J, Geleijnse Marcel L

机构信息

Department of Cardiology, The Thoraxcenter, Erasmus University Medical Center, 's-Gravendijkwal 230 Room BA 302, 3015 CE Rotterdam, The Netherlands.

出版信息

Eur J Echocardiogr. 2009 Jul;10(5):669-76. doi: 10.1093/ejechocard/jep036. Epub 2009 Apr 21.

Abstract

AIMS

This study sought to find the most robust method for left ventricular (LV) rotation measurement by speckle tracking echocardiography (STE) with the new QLAB Advanced Quantification Software (version 6.0, Philips, Best, The Netherlands).

METHODS AND RESULTS

The study population consisted of 40 non-selected patients (mean age 48 +/- 18 year, 20 men) and 50 non-selected healthy volunteers (mean age 34 +/- 12 year, 21 men). Feasibility and intra-observer reproducibility of the measurement of LV rotation parameters by STE were assessed for two different methods (Method A: six tracking points placed mid-myocardial and Method B: six tracking points placed endocardial and epicardial forming six myocardial segments). Subsequently, inter-observer and temporal reproducibility of the most robust method were assessed. Complete LV rotation assessment was more feasible with Method A (60 out of 90 subjects, 67% vs. 50 out of 90 subjects, 56%). In the 49 subjects in whom both Methods A and B were feasible, intra-observer reproducibility of LV rotation parameters was better with Method A (variabilities 2 +/- 3 to 10 +/- 9% vs. 2 +/- 4 to 21 +/- 18%). With this method, inter-observer variability varied from 4 +/- 4 to 13 +/- 9% and temporal variability from 4 +/- 6 to 19 +/- 15%.

CONCLUSION

The most robust method to assess LV rotation with QLAB software is from the mid-myocardium. This method is feasible in approximately two-thirds of subjects and has good intra-observer, inter-observer, and temporal reproducibility, allowing to study changes over time in LV rotation in an individual patient.

摘要

目的

本研究旨在利用新的QLAB高级定量软件(版本6.0,飞利浦,荷兰贝斯特),通过斑点追踪超声心动图(STE)找到测量左心室(LV)旋转的最可靠方法。

方法与结果

研究人群包括40例未经过挑选的患者(平均年龄48±18岁,男性20例)和50例未经过挑选的健康志愿者(平均年龄34±12岁,男性21例)。通过STE测量LV旋转参数的可行性和观察者内重复性,针对两种不同方法进行了评估(方法A:在心肌中层放置六个追踪点;方法B:在心内膜和心外膜放置六个追踪点,形成六个心肌节段)。随后,评估了最可靠方法的观察者间和时间重复性。使用方法A进行完整的LV旋转评估更可行(90名受试者中有60名,67%,而90名受试者中有50名,56%)。在方法A和方法B均可行的49名受试者中,方法A的LV旋转参数观察者内重复性更好(变异性为2±3至10±9%,而方法B为2±4至21±18%)。采用此方法,观察者间变异性为4±4至13±9%,时间变异性为4±6至19±15%。

结论

使用QLAB软件评估LV旋转的最可靠方法是从心肌中层进行测量。该方法在大约三分之二的受试者中可行,具有良好的观察者内、观察者间和时间重复性,能够研究个体患者LV旋转随时间的变化。