Davis T S, Torab K, House P, Greger B
Department of Bioengineering, University of Utah, Salt Lake City, UT 84112-9458, USA.
J Neurosci Methods. 2009 Jun 30;181(1):106-10. doi: 10.1016/j.jneumeth.2009.04.012. Epub 2009 Apr 24.
We have designed a device for long-term head fixation for use in behaving nonhuman primates that is robust yet minimally invasive and simple to use. This device is a modified version of the halo system that is used in humans for cervical traction and stabilization after spinal column injuries. This device consists of an aluminum halo with four titanium skull pins offset from the halo by aluminum posts. The titanium pins insert onto small segments of cranially reinforcing titanium plate, which are attached to the skull with titanium cortex screws. The surgery involves four scalp incisions, placement of the reinforcing plates, insertion of the pins for attachment of the halo, and incision closure. After the halo is attached, the animal's head can be fixed to a primate chair using a custom-built attachment arm that provides three degrees of adjustability for proper positioning during behavioral tasks. We have installed this device on two Macaque monkeys weighing 7 and 10kg. The halos have been in place on these animals for up to 8 months without signs of discomfort or loss of fixation. Using this method of head fixation, we have been able to track the animals' eye positions with an accuracy of less than two visual degrees while they perform behavioral tasks.
我们设计了一种用于行为学研究中的非人灵长类动物的长期头部固定装置,该装置坚固耐用、微创且易于使用。此装置是人类用于脊柱损伤后颈椎牵引和固定的halo系统的改良版本。该装置由一个铝制halo和四个钛制颅骨销组成,钛制颅骨销通过铝柱与halo偏移。钛制销插入颅骨强化钛板的小片段上,这些钛板用钛皮质螺钉固定在颅骨上。手术包括四个头皮切口、放置强化板、插入用于连接halo的销以及缝合切口。在连接halo后,动物的头部可以使用定制的连接臂固定在灵长类动物椅上,该连接臂在行为任务期间提供三个角度的可调节性以实现正确定位。我们已将此装置安装在两只体重分别为7千克和10千克的猕猴身上。这些动物身上的halo已放置长达8个月,没有不适或固定失效的迹象。使用这种头部固定方法,我们能够在动物执行行为任务时以小于两个视觉度的精度跟踪它们的眼睛位置。