Doheny Emer P, Lowery Madeleine M, O'Malley Mark J, Fitzpatrick David P
University College Dublin, Ireland.
Med Biol Eng Comput. 2009 Jun;47(6):589-98. doi: 10.1007/s11517-009-0488-2. Epub 2009 Apr 28.
Joint centre displacement may occur following total elbow replacement due to aseptic loosening or surgical misalignment, and has been linked to implant failure. In this study, the effects of joint centre displacement were examined using a neuromusculoskeletal model of the elbow joint. Isometric contractions were simulated at a range of joint angles during elbow flexion and extension. Displacement of the joint centre affected the force-generating capacity about the joint, due to changes in both muscle lengths and moment arms. The magnitude and direction of the maximum joint reaction force were also altered, potentially contributing to aseptic loosening and compromising joint stability. The relationship between force generated and the level of neural excitation to the elbow flexor and extensor muscles was also affected, suggesting that altered neural control patterns could be required following joint centre displacement.
全肘关节置换后,由于无菌性松动或手术对位不良可能会发生关节中心移位,这与植入物失败有关。在本研究中,使用肘关节的神经肌肉骨骼模型研究了关节中心移位的影响。在肘关节屈伸过程中的一系列关节角度下模拟了等长收缩。关节中心的移位影响了关节周围的力产生能力,这是由于肌肉长度和力臂的变化所致。最大关节反应力的大小和方向也发生了改变,这可能导致无菌性松动并损害关节稳定性。产生的力与肘关节屈肌和伸肌的神经兴奋水平之间的关系也受到了影响,这表明关节中心移位后可能需要改变神经控制模式。