Metallic artifacts caused by dental metal prostheses on PET images: a PET/CT phantom study using different PET/CT scanners.

作者信息

Shimamoto Hiroaki, Kakimoto Naoya, Fujino Kouichi, Hamada Seiki, Shimosegawa Eku, Murakami Shumei, Furukawa Souhei, Hatazawa Jun

机构信息

Department of Oral and Maxillofacial Radiology, Osaka University Graduate School of Dentistry, Suita, Osaka, Japan.

出版信息

Ann Nucl Med. 2009 Jul;23(5):443-9. doi: 10.1007/s12149-009-0254-4. Epub 2009 Apr 28.

Abstract

OBJECTIVE

The objective of this study was to investigate the effects of computed tomography (CT) artifacts caused by dental metal prostheses on positron emission tomography (PET) images.

METHODS

A dental arch cast was fixed in a cylindrical water-bath phantom. A spherical phantom positioned in the vicinity of the dental arch cast was used to simulate a tumor. To simulate the tumor imaging, the ratio of the (18)F-fluoro-deoxy-glucose radioactivity concentration of the spherical phantom to that of the water-bath phantom was set at 2.5. A dental bridge composed of a gold-silver-palladium alloy on the right mandibular side was prepared. A spherical phantom was set in the white artifact area on the CT images (site A), in a slightly remote area from the white artifact (site B), and in a black artifact area (site C). A PET/CT scan was performed with and without the metal bridge at each simulated tumor site, and the artifactual influence was evaluated on the axial attenuation-corrected (AC) PET images, in which the simulated tumor produced the strongest accumulation. Measurements were performed using three types of PET/CT scanners (scanners 1 and 2 with CT-based attenuation correction, and 3 with Cesium-137 ((137)Cs)-based attenuation correction). The influence of the metal bridge was evaluated using the change rate of the SUVmean with and without the metal bridge.

RESULTS

At site A, an overestimation was shown (scanner 1: +5.0% and scanner 2: +2.5%), while scanner 3 showed an underestimation of -31.8%. At site B, an overestimation was shown (scanner 1: +2.1% and scanner 2: +2.0%), while scanner 3 showed an underestimation of -2.6%. However, at site C, an underestimation was shown (scanner 1: -25.0%, scanner 2: -32.4%, and scanner 3: -8.4%).

CONCLUSIONS

When CT is used for attenuation correction in patients with dental metal prostheses, an underestimation of radioactivity of accumulated tracer is anticipated in the dark streak artifact area on the CT images. In this study, the dark streak artifacts of the CT caused by metallic dental prostheses may cause false negative finding of PET/CT in detecting small and/or low uptake tumor in the oral cavity.

摘要

目的

本研究旨在调查牙科金属假体引起的计算机断层扫描(CT)伪影对正电子发射断层扫描(PET)图像的影响。

方法

将牙弓模型固定在圆柱形水浴体模中。使用放置在牙弓模型附近的球形体模来模拟肿瘤。为模拟肿瘤成像,将球形体模的(18)F-氟脱氧葡萄糖放射性浓度与水浴体模的该浓度之比设定为2.5。制备右侧下颌由金银钯合金制成的牙桥。将球形体模置于CT图像上的白色伪影区域(A点)、距白色伪影稍远的区域(B点)和黑色伪影区域(C点)。在每个模拟肿瘤部位,分别在有和没有金属桥的情况下进行PET/CT扫描,并在轴向衰减校正(AC)PET图像上评估伪影影响,其中模拟肿瘤产生最强的放射性聚集。使用三种类型的PET/CT扫描仪进行测量(扫描仪1和2采用基于CT的衰减校正,扫描仪3采用基于铯-137((137)Cs)的衰减校正)。使用有和没有金属桥时SUVmean的变化率来评估金属桥的影响。

结果

在A点,出现高估(扫描仪1:+5.0%,扫描仪2:+2.5%),而扫描仪3显示低估为-31.8%。在B点,出现高估(扫描仪1:+2.1%,扫描仪2:+2.0%),而扫描仪3显示低估为-2.6%。然而,在C点,出现低估(扫描仪1:-25.0%,扫描仪2:-32.4%,扫描仪3:-8.4%)。

结论

当对有牙科金属假体患者使用CT进行衰减校正时,预计在CT图像上的暗条纹伪影区域会出现示踪剂聚集放射性的低估。在本研究中,牙科金属假体引起的CT暗条纹伪影可能导致PET/CT在检测口腔内小的和/或低摄取肿瘤时出现假阴性结果。