Nolan M F
Department of Anatomy, College of Medicine, University of South Florida, Tampa 33612.
Phys Ther. 1991 Oct;71(10):746-51. doi: 10.1093/ptj/71.10.746.
The purpose of this study was to document conductive differences among commercially available electrodes used with transcutaneous electrical nerve stimulation (TENS) devices. Impedance within a model system involving a human subject was calculated from oscilloscopic tracings of the pulse waveform for each of 25 different electrode types. Impedance values ranged from 1,000 to 7,800 omega Possible reasons for these differences are discussed. The observation that electrodes vary in their impedance and can thereby affect the stimulus applied to the skin raises the question of whether electrode choice might affect the clinical effectiveness of TENS. Attention is drawn to the skin electrodes as a variable that may affect the results of clinical and basic studies involving TENS.