Wong Catherine C L, Cociorva Daniel, Venable John D, Xu Tao, Yates John R
Department of Chemical Physiology, SR-11, The Scripps Research Institute, La Jolla, California 92037, USA.
J Am Soc Mass Spectrom. 2009 Aug;20(8):1405-14. doi: 10.1016/j.jasms.2009.04.007. Epub 2009 Apr 18.
We evaluate the effect of ion-abundance threshold settings for data-dependent acquisition on a hybrid LTQ-Orbitrap mass spectrometer, analyzing features such as the total number of spectra collected, the signal to noise ratio of the full MS scans, the spectral quality of the tandem mass spectra acquired, and the number of peptides and proteins identified from a complex mixture. We find that increasing the threshold for data-dependent acquisition generally decreases the quantity but increases the quality of the spectra acquired. This is especially true when the threshold setting is set above the noise level of the full MS scan. We compare two distinct experimental configurations: one where full MS scans are acquired in the Orbitrap analyzer while tandem MS scans are acquired in the LTQ analyzer, and one where both full MS and tandem MS scans are acquired in the LTQ analyzer. We examine the number of spectra, peptides, and proteins identified under various threshold conditions, and we find that the optimal threshold setting is at or below the respective noise level of the instrument regardless of whether the full MS scan is performed in the Orbitrap or in the LTQ analyzer. When comparing the high-throughput identification performance of the two analyzers, we conclude that, used at optimal threshold levels, the LTQ and the Orbitrap identify similar numbers of peptides and proteins. The higher scan speed of the LTQ, which results in more spectra being collected, is roughly compensated by the higher mass accuracy of the Orbitrap, which results in improved database searching and peptide validation software performance.
我们评估了数据依赖型采集的离子丰度阈值设置对混合LTQ-轨道阱质谱仪的影响,分析了诸如采集的谱图总数、全扫描质谱的信噪比、采集的串联质谱的谱图质量以及从复杂混合物中鉴定出的肽段和蛋白质数量等特征。我们发现,提高数据依赖型采集的阈值通常会减少采集到的谱图数量,但会提高其质量。当阈值设置高于全扫描质谱的噪声水平时,情况尤其如此。我们比较了两种不同的实验配置:一种是在轨道阱分析仪中进行全扫描质谱采集,而在LTQ分析仪中进行串联质谱采集;另一种是全扫描质谱和串联质谱均在LTQ分析仪中采集。我们研究了在各种阈值条件下鉴定出的谱图、肽段和蛋白质的数量,发现无论全扫描质谱是在轨道阱分析仪还是在LTQ分析仪中进行,最佳阈值设置都在仪器各自的噪声水平或以下。在比较两种分析仪的高通量鉴定性能时,我们得出结论,在最佳阈值水平下使用时,LTQ和轨道阱鉴定出的肽段和蛋白质数量相似。LTQ较高的扫描速度导致采集到更多的谱图,而轨道阱较高的质量精度则大致弥补了这一点,这使得数据库搜索和肽段验证软件的性能得到改善。