Suppr超能文献

经皮三尖瓣植入用一体式带瓣心包支架的动物可行性研究。

An integrated pericardial valved stent special for percutaneous tricuspid implantation: an animal feasibility study.

机构信息

Department of Cardiology, Changhai Hospital, the Second Military Medical University, Shanghai, PR China.

出版信息

J Surg Res. 2010 May 15;160(2):215-21. doi: 10.1016/j.jss.2008.10.029. Epub 2008 Nov 29.

Abstract

BACKGROUND

Various percutaneous valve replacement approaches have been reported in animals to replace the aortic and pulmonary valve. To broaden the indications of percutaneous approach to atrioventricular valves replacement, we developed a novel valved stent and evaluated the feasibility and safety of percutaneous implantation of the device in the tricuspid position.

MATERIALS AND METHODS

A unidirectional semilunar valve of porcine pericardium was sutured to a valvular ring. Then the ring with pericardial valve was mounted on a double-edge nitinol stent to construct the tricuspid valved stent. Transcatheter tricuspid valved stent implantation was performed on 10 healthy sheep. These sheep were followed up shortly after procedure with echocardiography evaluation and 64-slice CT imaging examination during the periodical follow-up at 1 mo and at 6 mo post-implantation. Additionally, two sheep were sacrificed after the procedure for anatomic and histological evaluation one at 1 h and the other at 1 mo, respectively.

RESULTS

Percutaneous valve implantation was successful in eight of 10 sheep. Two sheep died during the procedure due to migration of stent and fatal arrhythmia. The pressure of right heart did not significantly change after the procedure. Further echocardiography and imaging confirmed the stents were in desired position during the follow-up. The remaining six sheep with normal valvular and cardiac functionality survived for 6 mo after implantation.

CONCLUSIONS

The tricuspid stent with a valvular ring and pericardial valve can be implanted in tricuspid annulus percutaneous. The double-edge stent could substitute the native tricuspid valve chronically.

摘要

背景

各种经皮瓣膜置换方法已在动物中报道用于置换主动脉瓣和肺动脉瓣。为了拓宽经皮方法置换房室瓣的适应证,我们开发了一种新型带瓣支架,并评估了在三尖瓣位置经皮植入该装置的可行性和安全性。

材料和方法

将猪心包的单向半月形瓣膜缝合到瓣环上。然后将带瓣心包的环安装在双边缘镍钛诺支架上,构建三尖瓣带瓣支架。在 10 只健康绵羊上进行经导管三尖瓣带瓣支架植入。在手术后短期内通过超声心动图评估和 1 个月和植入后 6 个月的定期随访进行 64 层 CT 成像检查,对这些绵羊进行随访。此外,在手术后,两只绵羊分别在 1 小时和 1 个月时被处死,用于解剖和组织学评估。

结果

在 10 只绵羊中有 8 只成功进行了经皮瓣膜植入。由于支架移位和致命性心律失常,有 2 只绵羊在手术过程中死亡。手术后右心压力没有明显变化。进一步的超声心动图和影像学检查在随访期间确认支架处于理想位置。其余 6 只具有正常瓣膜和心脏功能的绵羊在植入后存活了 6 个月。

结论

带瓣环和心包瓣的三尖瓣支架可以经皮植入三尖瓣环。双边缘支架可以长期替代天然三尖瓣。

文献AI研究员