Serviço de Cabeça e Pescoço,Faculdade de Ciências Médicas, Universidade Estadual de Campinas.
Braz J Otorhinolaryngol. 2009 Jan-Feb;75(1):107-10. doi: 10.1016/s1808-8694(15)30840-5.
Tracheotomy is performed in cases of upper airway obstruction or chronic pulmonary disorders. The Tracheotomy Speech Valves (TSV) improve communication and airway hygiene and humidification of tracheotomized patients.
To show the low cost Brazilian TSV and its use in speech rehabilitation of tracheotomized patients, to evaluate diaphragm opening resistance and comfort to the patient.
Experimental, contemporary cohort.
The TSV was used in 32 patients. The valve has a diaphragm within a stainless steel body with plastic fittings. We studied the level of respiratory comfort according to the degree of valve diaphragm resistance, 40, 50 and 60 shores.
All the patients used the TSV coupled to the cannula in a regular basis, 26 of them did it for more than 12 hours daily and from these, 14 used it for 24h daily. The diaphragm pressure obtained was that of 40 shores for 13 patients and 50 shores for 19 patients. 60 shores was never used.
the metal TSV helps with speech without the need for closing the cannula with one's finger, and breathing was comfortable. We achieved standard diaphragm resistance. Currently all the patients from this study use this TSV with speech and 43.75% use it full time.
展示低成本的巴西气管造口术阀及其在气管造口患者言语康复中的应用,评估膈膜开启阻力和患者舒适度。
实验性、当代队列研究。
32 名患者使用了气管造口术阀。该阀有一个在不锈钢主体内的膈膜,带有塑料配件。我们根据阀膈膜阻力的程度来研究呼吸舒适度,分别为 40、50 和 60 肖氏硬度。
所有患者都定期将气管造口术阀与套管一起使用,其中 26 名患者每天使用超过 12 小时,其中 14 名患者每天使用 24 小时。获得的膈膜压力为 13 名患者为 40 肖氏硬度,19 名患者为 50 肖氏硬度。从未使用过 60 肖氏硬度。
金属气管造口术阀有助于言语,无需用手指关闭套管,呼吸也很舒适。我们达到了标准的膈膜阻力。目前,该研究中的所有患者都在使用这种 TSV 进行言语交流,其中 43.75%的患者全天使用。