Comparison of a microporous thermoplastic polyurethane-covered stent with a self-expanding bare nitinol stent in a porcine iliac artery model.

作者信息

Radeleff Boris, Grenacher Lars, Christoph Patricia, Sommer Christof M, Stampfl Ulrike, Ramsauer Stefanie, Henn Tina, Kurz Patrick, Lopez-Benitez Ruben, Berger Irina, Richter Goetz M

机构信息

Department of Diagnostic Radiology, University of Heidelberg, Im Neuenheimer Feld 110, 69120 Heidelberg, Germany.

出版信息

J Vasc Interv Radiol. 2009 Jul;20(7):927-35. doi: 10.1016/j.jvir.2009.04.044. Epub 2009 Jun 4.

Abstract

PURPOSE

To compare outcomes with a thermoplastic polyurethane (TPU)-covered self-expanding nitinol stent-graft (TPU graft) with those of a bare self-expanding nitinol stent in a porcine model.

MATERIALS AND METHODS

Fourteen TPU grafts and 14 commercially available bare nitinol stents were implanted, one each, in the iliac arteries of 14 minipigs. Follow-up was performed at 1 week (six animals), 4 weeks (four animals), and 12 weeks (four animals). The primary study endpoint was in-stent stenosis assessed with quantitative angiography and microscopy. Secondary endpoints were injury, inflammation, and endothelialization.

RESULTS

After 1 week, the maximum percentage luminal loss was significantly greater in TPU grafts (average, 16.2%; range, 0.0%-35.8%) than in bare nitinol stents (8.2%; 0.0%-17.3%) (P = .04). Three of the four TPU grafts were occluded after 4 weeks, and all four TPU grafts were occluded after 12 weeks. Binary stenosis was seen in three of four bare nitinol stents after both 4 and 12 weeks. At 4-week follow-up, the average percentage luminal loss was significantly greater in TPU grafts (85.2%; 40.8%-100%) than in bare nitinol stents (49.5%; 37.9%-62.4%) (P = .003). The difference in neointimal height and percentage average stenosis between TPU grafts (1,028.7 microm and 68.4%) and bare nitinol stents (1,033.6 microm [918.0-1,118.40 microm] and 68.1% [60.44%-71.99%]) was not statistically significant. After 12 weeks, the average percentage luminal loss was 100% in TPU grafts due to occlusion of all stent-grafts and 24.9% (8.0%-63.9%) in bare nitinol stents (P = .011).

CONCLUSIONS

TPU grafts failed to provide improved patency compared with the bare nitinol stents because of excessive neointimal growth and subsequent occlusion. In addition, the bare nitinol stents showed considerable in-stent stenosis at angiography and microscopy.

摘要

目的

在猪模型中比较热塑性聚氨酯(TPU)覆盖的自膨式镍钛诺支架移植物(TPU移植物)与裸自膨式镍钛诺支架的效果。

材料与方法

将14个TPU移植物和14个市售裸镍钛诺支架分别植入14只小型猪的髂动脉。在1周(6只动物)、4周(4只动物)和12周(4只动物)进行随访。主要研究终点是通过定量血管造影和显微镜评估的支架内狭窄。次要终点是损伤、炎症和内皮化。

结果

1周后,TPU移植物的最大管腔损失百分比(平均显著(平均16.2%;范围0.0%-35.8%)显著高于裸镍钛诺支架(8.2%;0.0%-17.3%)(P = 0.04)。4周后,4个TPU移植物中有3个闭塞,12周后所有4个TPU移植物均闭塞。4周和12周后,4个裸镍钛诺支架中有3个出现二元狭窄。在4周随访时,TPU移植物的平均管腔损失百分比(85.2%;40.8%-100%)显著高于裸镍钛诺支架(49.5%;37.9%-62.4%)(P = 0.003)。TPU移植物(1028.7微米和68.4%)与裸镍钛诺支架(1033.6微米[918.0-1118.40微米]和68.1%[60.44%-71.99%])之间的新生内膜高度和平均狭窄百分比差异无统计学意义。12周后,由于所有支架移植物闭塞,TPU移植物的平均管腔损失百分比为100%,裸镍钛诺支架为24.9%(8.0%-63.9%)(P = 0.011)。

结论

由于过度的新生内膜生长和随后的闭塞,TPU移植物与裸镍钛诺支架相比未能提供更好的通畅性。此外,裸镍钛诺支架在血管造影和显微镜检查中显示出相当程度的支架内狭窄。