Evaluation of the effectiveness of a wearable cardioverter defibrillator detection algorithm.

作者信息

Dillon Katie A, Szymkiewicz Steven J, Kaib Thomas E

机构信息

ZOLL Lifecor Corporation, Pittsburgh, PA, USA.

出版信息

J Electrocardiol. 2010 Jan-Feb;43(1):63-7. doi: 10.1016/j.jelectrocard.2009.05.010.

Abstract

BACKGROUND

Wearable cardioverter defibrillators (WCDs) provide protection from sudden cardiac death. The efficacy of a WCD detection algorithm has not been reported outside of clinical trial.

METHODS

The efficacy of the algorithm was reviewed through a retrospective analysis of appropriate shocks, inappropriate shocks, and arrhythmia detections during a 1-year period.

RESULTS

WCD patients had an appropriate shock rate of 1.58 per 100 patient-months and an inappropriate shock rate of 0.99 per 100 patient-months. Most of the arrhythmia detections in a 3-month period were short in length, with only 2.7% of the detections lasting over 25 seconds, the time at which a shock becomes possible.

CONCLUSIONS

By incorporating a patient responsiveness test, as well as features that eliminate or reduce signal interference common to external electrocardiogram electrodes, the WCD detection algorithm has a low risk of inappropriate shocks.