Morte Elane S Boa, Korn Maria Graças A, Saraiva M Lúcia M F S, Lima José L F C, Pinto Paula C A G
PRONEX - Instituto de Química, Universidade Federal da Bahia, Salvador, Bahia, Brazil.
Talanta. 2009 Sep 15;79(4):1100-3. doi: 10.1016/j.talanta.2009.02.019. Epub 2009 Feb 20.
The present work describes the development of a fast and robust sequential injection fluorimetric procedure for the determination of Sn in juices of canned fruits. The developed automatic methodology is based on the complexation of Sn with 8-hydroxyquinoline-5-sulfonic acid (HQSA) to form a fluorimetric product (lambda(exc)=354 nm; lambda(em)=510 nm). The influence of dimethylsulfoxide (DMSO) and cetylpyridinium bromide (CPB) on the sensitivity of the fluorimetric determination was evaluated. Linear calibration plots were obtained for Sn concentrations between 1 and 10 mg L(-1), with a detection limit of 0.38 mg L(-1). In each analytical cycle 0.006 mg of HQSA and 0.47 mg of CPB were consumed and 1.5 mL of effluent was generated. The developed methodology was applied to the determination of Sn in juices of canned fruits and the results complied with those furnished by an electrothermal atomic absorption spectrometry comparison procedure, with relative deviations lower than 5.2%. The automatic procedure exhibited good precision (R.S.D.<1.4%) and the sampling rate was about 70 determinations per hour.
本研究描述了一种快速且稳健的顺序注射荧光法的开发,用于测定罐装水果汁中的锡。所开发的自动方法基于锡与8-羟基喹啉-5-磺酸(HQSA)络合形成荧光产物(激发波长λ(exc)=354 nm;发射波长λ(em)=510 nm)。评估了二甲基亚砜(DMSO)和十六烷基溴化吡啶(CPB)对荧光测定灵敏度的影响。获得了锡浓度在1至10 mg L(-1)之间的线性校准曲线,检测限为0.38 mg L(-1)。在每个分析周期中,消耗0.006 mg的HQSA和0.47 mg的CPB,并产生1.5 mL的流出物。所开发的方法应用于罐装水果汁中锡的测定,结果与电热原子吸收光谱比较法提供的结果相符,相对偏差低于5.2%。该自动方法具有良好的精密度(相对标准偏差R.S.D.<1.4%),采样率约为每小时70次测定。