Department of Orthopaedics, Radboud University, Nijmegen Medical Centre, The Netherlands.
Acta Orthop. 2009 Aug;80(4):440-4. doi: 10.3109/17453670903153568.
The first generations of total ankle replacements (TARs) showed a high rate of early failure. In the last decades, much progress has been made in the development of TARs, with the newer generation showing better results. We evaluated TARs implanted with rheumatoid arthritis (RA) or juvenile inflammatory arthritis (JIA) as indication.
58 total ankle prostheses (Buechel-Pappas and STAR type) were implanted in patients with RA (n = 53) or JIA (n = 5) in 54 patients (4 bilateral). After a mean followup of 2.7 (1-9) years, all patients were reviewed by two orthopedic surgeons who were not the surgeons who performed the operation. Standard AP and lateral radiographs were taken and a Kofoed ankle score was obtained; this is a clinical score ranging from 0-100 and consists of sub-scores for pain, disability, and range of motion.
2 patients died of unrelated causes. Of the 52 patients who were alive (56 prostheses), 51 implants were still in place and showed no signs of loosening on the most recent radiographs. The mean Kofoed score at follow-up was 73 points (SD 16, range 21-92). 4 patients showed a poor result (score < 50) with persistent pain for which no obvious reason could be found. 5 implants were removed, 4 because of infection and 1 because of aseptic loosening.
Medium-term results of the STAR and BP types of TAR in RA were satisfactory. The main reason for failure of the implant was infection.
第一代全踝关节置换术(TAR)早期失败率较高。在过去的几十年中,TAR 的发展取得了很大的进展,新一代产品的效果更好。我们评估了类风湿关节炎(RA)或青少年特发性关节炎(JIA)作为适应证的 TAR。
54 例患者(4 例双侧)中,53 例 RA 患者和 5 例 JIA 患者共植入 58 例全踝关节假体(Buechel-Pappas 和 STAR 型)。平均随访 2.7(1-9)年后,由两位未行手术的矫形外科医生对所有患者进行了复查。拍摄了标准的前后位和侧位 X 线片,并获得了 Kofoed 踝关节评分;这是一个临床评分,范围从 0 到 100 分,包括疼痛、残疾和活动范围的子评分。
2 例患者因非相关原因死亡。在 52 例存活的患者(56 例假体)中,51 例假体仍在位,最近的 X 线片未见松动迹象。随访时的平均 Kofoed 评分为 73 分(SD 16,范围 21-92)。4 例患者出现持续疼痛但评分<50,原因不明,结果较差。5 例假体被取出,4 例因感染,1 例因无菌性松动。
RA 患者中 STAR 和 BP 型 TAR 的中期结果令人满意。植入物失败的主要原因是感染。