Shiehmorteza Masoud, Sirlin Claude B, Wolfson Tanya, Gamst Anthony, Soumekh Andre E, Heaton Shay P, Kohl Chad A, Pinto Nicholas, Bydder Mark
Department of Radiology, University of California, San Diego, California, USA.
J Magn Reson Imaging. 2009 Sep;30(3):547-53. doi: 10.1002/jmri.21861.
To show the signal intensity varies with shot number in diffusion-weighted (DW) echo-planar imaging (EPI) and affects apparent diffusion coefficient (ADC) calculation.
This prospective study was performed on 35 adult patients and 20 volunteers. Measurements were made on a 3T scanner using a breathhold DW spin-echo EPI (SE EPI) sequence. Three protocols were used: A) eight consecutive shots at a fixed b-value of 0 seconds/mm(2) with TR = 1000 and 3000 msec; B) seven consecutive shots at b-values = 0, 1000, 750, 500, 250, 100, 0 seconds/mm(2) (in that order) with TR = 3500 msec; and C) seven consecutive shots (as in B) with TR = 1000, 1750, and 7000 msec.
For protocol A, signal intensity decreased significantly from the first to second shot (P<0.0001) and thereafter remained constant. For protocol B, the ADC depended on which b = 0 seconds/mm(2) image was used. Using the first b = 0 seconds/mm(2), the mean ADC was 15% higher than using the second b = 0 seconds/mm(2) (P<0.0001). For protocol C, the difference between ADC using the first b = 0 seconds/mm(2) and the second b = 0 seconds/mm(2) decreased as the TR increased.
The signal intensity can vary with shot number in SE EPI. For TR > or = 3000 msec, steady-state is attained after one shot. Using data acquired prior to steady-state confounds the calculation of ADC values.
展示在扩散加权(DW)回波平面成像(EPI)中信号强度随激发次数变化,并影响表观扩散系数(ADC)的计算。
对35名成年患者和20名志愿者进行了这项前瞻性研究。在3T扫描仪上使用屏气DW自旋回波EPI(SE EPI)序列进行测量。使用了三种方案:A)在固定b值为0秒/毫米²(TR = 1000和3000毫秒)下连续进行8次激发;B)在b值 = 0、1000、750、500、250、100、0秒/毫米²(按此顺序)下连续进行7次激发,TR = 3500毫秒;C)与B相同连续进行7次激发,TR = 1000、1750和7000毫秒。
对于方案A,信号强度从第一次激发到第二次激发显著降低(P<0.0001),此后保持恒定。对于方案B,ADC取决于使用的哪个b = 0秒/毫米²图像。使用第一个b = 0秒/毫米²时,平均ADC比使用第二个b = 0秒/毫米²时高15%(P<0.0001)。对于方案C,随着TR增加,使用第一个b = 0秒/毫米²和第二个b = 0秒/毫米²时ADC的差异减小。
在SE EPI中信号强度会随激发次数变化。对于TR≥3000毫秒,一次激发后达到稳态。使用稳态之前采集的数据会混淆ADC值的计算。