Rudolph Ross, Forcier Nancy
University of California San Diego, San Diego, CA, USA.
Aesthet Surg J. 2009 Jul-Aug;29(4):310-3. doi: 10.1016/j.asj.2009.02.009.
Patients needing silicone gel-filled breast implant surgical revision often do not have information as to the size of their implants. Clinical estimates may be erroneous and an accurate, in vivo method of determining breast implant volume would be very useful in planning surgery.
Our objective was to determine whether magnetic resonance imaging (MRI) scans of the breast in combination with a computer-assisted detection (CAD) system could provide accurate estimation of implant volumes in patients with silicone gel-filled implants, using known implant volumes for comparison.
All MRI scans were performed for medical necessity before implant revision surgery or for follow-up. Scans and silicone implant volume calculations were performed in 20 patients (39 breasts). Fifteen patients (n = 27 breasts) had known implant volumes. The MRI images were analyzed blindly by a single radiologist using the image processing features of an MRI CAD system (DynaCAD; Invivo, Birmingham, MI) in a novel way. Computed implant volumes were compared to known implant volumes obtained from documented preoperative notes (n = 17) or following implant removal via implant markings (n = 10).
Average deviation of the 27 known implant volumes from those calculated by MRI DynaCAD analysis was only 0.82% (SD, 3.95%). In four of 27 breasts (14.8%), MRI calculations matched the clinical volumes exactly. The widest volume deviation range for all implants was +11.4% to -4.44%. A variety of implant types were found, with 11 confirmed ruptures.
The calculation of silicone gel-filled breast implant volume using breast MRI scans and a commercially-available CAD system appears to be sufficiently accurate that it may have significant clinical benefit in planning revision implant surgery. Calculations can be easily obtained in five minutes. To our knowledge, this is the first report describing this method to reliably measure silicone implant volumes preoperatively.
需要进行硅胶填充乳房植入物手术修复的患者通常不知道其植入物的尺寸。临床估计可能有误,而一种准确的体内确定乳房植入物体积的方法对于手术规划将非常有用。
我们的目的是确定乳房磁共振成像(MRI)扫描结合计算机辅助检测(CAD)系统能否通过与已知植入物体积进行比较,为硅胶填充植入物患者提供准确的植入物体积估计。
所有MRI扫描均因医疗需要在植入物修复手术前或随访时进行。对20例患者(39个乳房)进行了扫描和硅胶植入物体积计算。15例患者(n = 27个乳房)已知植入物体积。由一名放射科医生以一种新颖的方式使用MRI CAD系统(DynaCAD;Invivo,伯明翰,密歇根州)的图像处理功能对MRI图像进行盲法分析。将计算出的植入物体积与从术前记录(n = 17)或通过植入物标记取出植入物后获得的已知植入物体积(n = 10)进行比较。
27个已知植入物体积与通过MRI DynaCAD分析计算出的体积的平均偏差仅为0.82%(标准差,3.95%)。在27个乳房中的4个(14.8%),MRI计算结果与临床体积完全匹配。所有植入物的最大体积偏差范围为+11.4%至 -4.44%。发现了多种植入物类型,其中11个确认破裂。
使用乳房MRI扫描和市售CAD系统计算硅胶填充乳房植入物体积似乎足够准确,可能在规划植入物修复手术中具有显著的临床益处。计算可在五分钟内轻松完成。据我们所知,这是第一份描述这种术前可靠测量硅胶植入物体积方法的报告。