Saldanha Mario J, Benjamin Larry, Patel Chetan K
Oxford Eye Hospital, UK.
J Cataract Refract Surg. 2009 Oct;35(10):1751-5. doi: 10.1016/j.jcrs.2009.05.024.
To assess the rotational stability of the AcrySof MA60BM acrylic intraocular lens (IOL) with and without aspiration of residual lens epithelial cells (LECs).
Stoke Mandeville Hospital, Aylesbury, United Kingdom.
Routine phacoemulsification was performed with (Group A) or without (Group B) aspiration of residual LECs (random allocation). The baseline position of the IOL was determined from a video frame acquired at the conclusion of surgery. Postoperative IOL position was documented using digital retroillumination images at 2 weeks and 3 months. Early IOL rotation (surgery to 2 weeks postoperatively) was graded as mild (<10 degrees), moderate (10 to 30 degrees), or severe (>30 degrees) by semiobjective comparison of the images. Late IOL rotation (2 weeks to 3 months) was measured more precisely using purpose-designed software.
Group A had incomplete aspiration of residual LECs. The only between-group demographic difference was significantly more men in Group A than in Group B (P<.05). There were no cases of severe early IOL rotation. Mild and moderate early rotation occurred in 91.5% of eyes and 8.5% of eyes, respectively. The mean late rotation was 0.03 degrees +/- 3.06 (SD) (range 7.8 to -8.7 degrees). Aspiration of residual LECs made no statistical difference in early or late rotation (P>.05). Between 2 weeks and 3 months, counterclockwise rotation occurred in 50% of all cases.
Postoperative rotation of the loop-haptic IOL was stable; LEC clearance did not adversely affect performance. The tendency toward counterclockwise rotation would not preclude good performance of a toric model of the IOL.
评估有或无残留晶状体上皮细胞(LEC)抽吸情况下,AcrySof MA60BM丙烯酸人工晶状体(IOL)的旋转稳定性。
英国艾尔斯伯里斯托克曼德维尔医院。
进行常规超声乳化手术,有(A组)或无(B组)残留LEC抽吸(随机分配)。IOL的基线位置由手术结束时获取的视频帧确定。术后2周和3个月使用数字后照光图像记录IOL位置。通过图像的半客观比较,将早期IOL旋转(手术至术后2周)分为轻度(<10度)、中度(10至30度)或重度(>30度)。使用专门设计的软件更精确地测量晚期IOL旋转(2周至3个月)。
A组残留LEC抽吸不完全。两组之间唯一的人口统计学差异是A组男性明显多于B组(P<.05)。没有严重早期IOL旋转的病例。轻度和中度早期旋转分别发生在91.5%的眼和8.5%的眼。晚期平均旋转为0.03度±3.06(标准差)(范围7.8至 -8.7度)。残留LEC抽吸在早期或晚期旋转方面无统计学差异(P>.05)。在2周和3个月之间,50%的所有病例出现逆时针旋转。
袢状触觉IOL术后旋转稳定;LEC清除对性能无不利影响。逆时针旋转的趋势不排除IOL复曲面模型的良好性能。