Wen Ying, Shan Guiqiu, Pu Jinhui, Liu Zhongming
Department of Blood Transfusion, General Hospital of Guangzhou Command, Guangzhou 510010, China.
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2009 Aug;26(4):883-5.
To research piezoelectric immunosensor array for rapid detecting HIV(1+2), piezoelectric immunosensor array matrix was designed. HIV(1+2)C1 antigen was immobilized onto the silver electrodes of quartz-crystal microbalance, which was modified with adsorption and cross-linked method. In the clean air flow and monitoring environment, standard quality control and clinical serum sample were detected. The linear range for the measurement of HIV(1+2) was 0.01-0.2 NCU/ml. The sensitivity, specificity and accuracy of HIV(1+2) piezoelectric protein sensor array were 91.7%, 93.3% and 92.7% respectively.