使用Amplatzer封堵器关闭室间隔缺损失败。

Failed closure of a ventricular septal defect with an Amplatzer occluder.

作者信息

Wippermann Jens, Hoppe Uta C, Sreeram Narayanswami, Wahlers Thorsten

机构信息

Department of Cardiothoracic Surgery, Heart Center, University Hospital of Cologne, Kerpener Strasse 62, 50937 Cologne, Germany.

出版信息

Interact Cardiovasc Thorac Surg. 2010 Feb;10(2):320-1. doi: 10.1510/icvts.2009.216200. Epub 2009 Nov 13.

Abstract

A 74-year-old man was diagnosed to have a ventricular septal defect (VSD), which was detected shortly following transvenous pacemaker implantation. Transoesophageal echocardiography suggested the presence of two VSDs, one of which was closed with a device. At surgery, a single large VSD was seen, with the implanted device having embolised into the left ventricle. The defect was successfully closed using a pericardial patch, and the embolised device explanted.

摘要

一名74岁男性在经静脉起搏器植入后不久被诊断出患有室间隔缺损(VSD)。经食管超声心动图显示存在两个室间隔缺损,其中一个已用装置封堵。手术中发现一个单一的大室间隔缺损,植入的装置已栓塞至左心室。使用心包补片成功闭合缺损,并取出栓塞的装置。