A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup.

机构信息

Department of Radiation Oncology, Good Samaritan Health System, Kearney, NE, USA.

Henry Ford Down River Center for Oncology, Trenton, MI, USA.

出版信息

J Appl Clin Med Phys. 2009 Oct 6;10(4):273-280. doi: 10.1120/jacmp.v10i4.2954.

Abstract

A positioning QA procedure for Varian's 2D/2D (kV/MV) and 3D/3D (planCT/CBCT) matching was developed. The procedure was to check: (1) the coincidence of on-board imager (OBI), portal imager (PI), and cone beam CT (CBCT)'s isocenters (digital graticules) to a linac's isocenter (to a pre-specified accuracy); (2) that the positioning difference detected by 2D/2D (kV/MV) and 3D/3D(planCT/CBCT) matching can be reliably transferred to couch motion. A cube phantom with a 2 mm metal ball (bb) at the center was used. The bb was used to define the isocenter. Two additional bbs were placed on two phantom surfaces in order to define a spatial location of 1.5 cm anterior, 1.5 cm inferior, and 1.5 cm right from the isocenter. An axial scan of the phantom was acquired from a multislice CT simulator. The phantom was set at the linac's isocenter (lasers); either AP MV/R Lat kV images or CBCT images were taken for 2D/2D or 3D/3D matching, respectively. For 2D/2D, the accuracy of each device's isocenter was obtained by checking the distance between the central bb and the digital graticule. Then the central bb in orthogonal DRRs was manually moved to overlay to the off-axis bbs in kV/MV images. For 3D/3D, CBCT was first matched to planCT to check the isocenter difference between the two CTs. Manual shifts were then made by moving CBCT such that the point defined by the two off-axis bbs overlay to the central bb in planCT. (PlanCT can not be moved in the current version of OBI1.4.) The manual shifts were then applied to remotely move the couch. The room laser was used to check the accuracy of the couch movement. For Trilogy (or Ix-21) linacs, the coincidence of imager and linac's isocenter was better than 1 mm (or 1.5 mm). The couch shift accuracy was better than 2 mm.

摘要

我们开发了一种用于瓦里安的二维/二维(kV/MV)和三维/三维(计划 CT/CBCT)配准的定位 QA 程序。该程序旨在检查:(1)机载成像仪(OBI)、门户成像仪(PI)和锥形束 CT(CBCT)的等中心(数字栅格)与直线加速器的等中心(达到预定精度)的一致性;(2)通过二维/二维(kV/MV)和三维/三维(计划 CT/CBCT)匹配检测到的定位差异能够可靠地转换为治疗床运动。使用带有中心 2mm 金属球(bb)的立方体形模。bb 用于定义等中心。在两个模体表面上放置另外两个 bb,以定义等中心前 1.5cm、下 1.5cm 和右 1.5cm 的空间位置。使用多排 CT 模拟器对模体进行轴向扫描。将模体设置在直线加速器的等中心(激光)处;分别拍摄 AP MV/R Lat kV 图像或 CBCT 图像,用于二维/二维或三维/三维匹配。对于二维/二维,通过检查中心 bb 与数字栅格之间的距离来获得每个设备等中心的准确性。然后手动将正交 DRR 中的中心 bb 移动以与 kV/MV 图像中的偏心 bb 重叠。对于三维/三维,首先将 CBCT 与计划 CT 匹配,以检查两个 CT 之间的等中心差异。然后通过移动 CBCT 使两个偏心 bb 定义的点与计划 CT 中的中心 bb 重叠来进行手动移位。(在当前版本的 OBI1.4 中,无法移动计划 CT。)然后将手动移位应用于远程移动治疗床。使用房间激光检查治疗床运动的准确性。对于 Trilogy(或 Ix-21)直线加速器,成像仪和直线加速器等中心的一致性优于 1mm(或 1.5mm)。治疗床移位精度优于 2mm。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/4837/5720565/6f746eb2b81f/ACM2-10-273-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验