使用射频电能经房间隔穿刺针的初步经验。

Initial experience using a radiofrequency powered transseptal needle.

机构信息

Section of Cardiology, Department of Internal Medicine, University of Chicago, Chicago, Illinois, USA.

出版信息

J Cardiovasc Electrophysiol. 2010 Apr;21(4):423-7. doi: 10.1111/j.1540-8167.2009.01656.x. Epub 2009 Nov 17.

Abstract

INTRODUCTION

The purpose of this study was to determine the safety and efficacy of using a novel radiofrequency (RF) powered transseptal needle to perform transseptal puncture (TSP).

METHODS

TSP was performed in 35 consecutive patients undergoing left-sided catheter ablation (mean age = 51 years; male = 71%) using a RF powered transseptal needle (NRG, Adult Large and Standard Curve C1, 71 cm, Baylis Medical Company, Inc.). Prior TSP had been performed in 34% of patients. The transseptal apparatus was positioned with the tip of the dilator engaged in the fossa ovalis. RF energy was delivered to the tip of the transseptal needle using a proprietary RF generator at 10 W for 2 seconds as gentle pressure was applied to the needle.

RESULTS

In 5 of the 41 TSPs, the needle crossed into the left atrium before RF energy was delivered. In 35 of the remaining 36 punctures, the needle was successfully advanced into the left atrium after application of RF current. In 1 patient, the TSP with the powered needle was unsuccessful but was accomplished using a standard needle. The only complication was a transient right atrial thrombus, which occurred in 2 patients.

CONCLUSION

A radiofrequency powered transseptal needle can be used to perform TSP safely and successfully without the need for significant mechanical force, even in patients who have undergone TSP previously. Additional studies are needed to determine whether a powered transseptal needle should be used routinely.

摘要

简介

本研究旨在确定使用新型射频(RF)动力经间隔穿刺针进行经间隔穿刺(TSP)的安全性和有效性。

方法

使用 RF 动力经间隔穿刺针(NRG,成人大型和标准曲线 C1,71cm,Baylis Medical Company,Inc.)对 35 例接受左侧导管消融的连续患者进行 TSP(平均年龄=51 岁;男性=71%)。在 34%的患者中已经进行了先前的 TSP。将经间隔器械的尖端置于卵圆窝中进行定位。使用专有 RF 发生器将 RF 能量施加到经间隔针的尖端,施加 2 秒 10W 的能量,同时对针施加轻柔的压力。

结果

在 41 次 TSP 中有 5 次,在施加 RF 能量之前,针已经穿过房间隔进入左心房。在剩余的 36 次穿刺中的 35 次中,在施加 RF 电流后,针成功地进入左心房。在 1 例患者中,使用动力针的 TSP 不成功,但使用标准针完成。唯一的并发症是 2 例患者出现短暂的右心房血栓。

结论

即使在先前已经进行 TSP 的患者中,也可以安全且成功地使用射频动力经间隔穿刺针进行 TSP,而无需施加显著的机械力。需要进一步的研究来确定是否应常规使用动力经间隔穿刺针。