Receptor affinity CE for measuring bioactive inflammatory cytokines in human skin biopsies.

机构信息

Laboratory of Bioengineering and Physical Science, National Institute of Biomedical Imaging and Bioengineering, National Institutes of Health, Bethesda, MD 20892, USA.

出版信息

Electrophoresis. 2009 Nov;30(22):3947-54. doi: 10.1002/elps.200900311.

Abstract

A chip-based receptor affinity CE system has been employed to measure the concentrations of bioactive pro-inflammatory cytokines in biopsy materials obtained from human atopic skin lesions. The device employs a replaceable affinity disk to which recombinant cytokine receptors have been chemically immobilized. Homogenates obtained from micro-dissected human skin samples were injected into the system where the bioactive cytokines were captured in the receptor affinity port and labeled in situ with a laser dye. The captured cytokines were released and separated by CE, the resolved peaks being detected and measured by laser-induced fluorescence. When compared with conventional cell-based bioassays, the affinity receptor chip showed reasonable correlation with r(2) values of 0.998 for interferon gamma, 0.994 for IL-6 and 0.991 for tumor necrosis factor alpha. The complete process including cytokine capture, labeling, and analysis took approximately 12.5 min with intra- and inter-assay CVs below 5.3% and recoveries of 84.9-98.4% at the 100 pg/mL concentration in buffer solutions and 84.5-95% in normal human tissue extract. The system could indicate clear differences between the various clinical stages of atopic dermatitis in human patients and could run 4-6 samples per hour. This system, like previous chip-based systems designed in our laboratory, holds the potential for being modified to be a portable unit that could be used in clinics and other biomedical screening studies.

摘要

基于芯片的受体亲和 CE 系统已被用于测量从人类特应性皮肤损伤活检材料中获得的生物活性促炎细胞因子的浓度。该装置采用可更换的亲和盘,盘上化学固定有重组细胞因子受体。从微解剖的人体皮肤样本中获得的匀浆被注入该系统,生物活性细胞因子在受体亲和端口被捕获,并在原位用激光染料标记。捕获的细胞因子通过 CE 释放和分离,解析峰通过激光诱导荧光进行检测和测量。与传统的基于细胞的生物测定相比,亲和受体芯片显示出合理的相关性,r(2) 值分别为干扰素γ 0.998、IL-6 0.994 和肿瘤坏死因子α 0.991。整个过程包括细胞因子的捕获、标记和分析,大约需要 12.5 分钟,在缓冲溶液中 100 pg/mL 浓度下的内和间测定 CV 低于 5.3%,回收率为 84.9-98.4%,在正常人组织提取物中的回收率为 84.5-95%。该系统可以清楚地区分人类特应性皮炎的不同临床阶段,每小时可以运行 4-6 个样本。该系统与我们实验室之前设计的基于芯片的系统类似,具有被修改为便携式设备的潜力,可用于临床和其他生物医学筛选研究。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验