Bharkhada Deepak, Yu Hengyong, Dixon Robert, Wei Yuchuan, Carr J Jeffrey, Bourland J Daniel, Best Ryan, Hogan Richard, Wang Ge
Biomedical Imaging Division, VT-WFU School of Biomedical Engineering & Science, Wake Forest University, Winston-Salem, NC 27157, USA.
J Comput Assist Tomogr. 2009 Nov-Dec;33(6):967-72. doi: 10.1097/RCT.0b013e31819f61e2.
With continuing developments in computed tomography (CT) technology and its increasing use of CT imaging, the ionizing radiation dose from CT is becoming a major public concern particularly for high-dose applications such as cardiac imaging. We recently proposed a novel interior tomography approach for x-ray dose reduction that is very different from all the previously proposed methods. Our method only uses the projection data for the rays passing through the desired region of interest. This method not only reduces x-ray dose but scatter as well. In this paper, we quantify the reduction in the amount of x-ray dose and scattered radiation that could be achieved using this method. Results indicate that interior tomography may reduce the x-ray dose by 18% to 58% and scatter to the detectors by 19% to 59% as the FOV is reduced from 50 to 8.6 cm.
随着计算机断层扫描(CT)技术的不断发展及其在CT成像中的应用日益增加,CT产生的电离辐射剂量正成为一个主要的公众关注点,尤其是对于心脏成像等高剂量应用。我们最近提出了一种用于降低X射线剂量的新型内部断层扫描方法,该方法与之前提出的所有方法都有很大不同。我们的方法仅使用穿过所需感兴趣区域的射线的投影数据。这种方法不仅能降低X射线剂量,还能减少散射。在本文中,我们对使用该方法可实现的X射线剂量和散射辐射量的减少进行了量化。结果表明,随着视野从50厘米减小到8.6厘米,内部断层扫描可将X射线剂量降低18%至58%,并将散射到探测器的剂量降低19%至59%。