T 波过感知和心脏再同步治疗中的双心室起搏比例低。

T wave oversensing and low percentage of biventricular pacing in cardiac resynchronization therapy.

机构信息

Hospital Donostia Arrythmia Unit, Intensive Care Department, Donostia, Spain.

出版信息

Cardiol J. 2009;16(6):580-1.

Abstract

The success of cardiac resynchronization therapy (CRT) depends on a high percentage of ventricular pacing. We present the electrocardiography and electrograms of a patient who underwent an implantable cardioverter-defibrillator (ICD) with CRT implantation showing a low percentage of ventricular pacing as a result of T wave oversensing of paced QRS. The patient showed no clinical improvement. We suggest finding a good sensitivity of F waves in induced ventricular fibrillation during ICD implantation in order to overcome the harm of possible oversensing.

摘要

心脏再同步治疗(CRT)的成功取决于较高的心室起搏比例。我们介绍了一位植入式心脏复律除颤器(ICD)联合 CRT 植入术患者的心电图和心内电图,结果显示由于起搏 QRS 波的 T 波过感知导致心室起搏比例较低。患者无临床改善。我们建议在 ICD 植入过程中寻找诱发室颤时 F 波的良好灵敏度,以克服可能过感知的危害。