Kluenter Heinz-Dieter, Lang-Roth Ruth, Beutner Dirk, Hüttenbrink Karl-Bernd, Guntinas-Lichius Orlando
Department of Otorhinolaryngology, Head and Neck Surgery, University of Cologne, Cologne, Germany.
Acta Otolaryngol. 2010 Jun;130(6):696-701. doi: 10.3109/00016480903373732.
The approach to the cochlea did not influence postural control after cochlear implantation (CI) surgery. Most patients already have impaired vestibular function before surgery. These balance deficits did not change after CI surgery in the majority of patients but static balance might be improved by CI surgery.
To determine whether the technique used to insert the CI electrode has an influence on vestibular function.
Static and dynamic postural control and caloric vestibular function were measured prospectively before and after CI surgery in 36 CI patients with standard cochleostomy (SC) insertion and 16 patients with insertion across the round window (RW). The test battery consisted of the modified clinical test of sensory interaction on balance (mCTSIB), the Rhythmic Weight Shift (RWS), the Walk Across (WA), and the Tandem Walk (TW) test.
Caloric testing was normal in the majority of CI candidates and remained normal after surgery. mCTSIB results were not significantly different before and after CI surgery between the SC and RW groups. Within both subgroups, the mCTSIB composite value improved within 6 weeks after surgery compared with the values obtained before surgery. Dynamic standing balance showed no postoperative change in RWS.
耳蜗植入(CI)手术后,进入耳蜗的方式对姿势控制没有影响。大多数患者在手术前就已经存在前庭功能受损的情况。在大多数患者中,这些平衡缺陷在CI手术后并未改变,但CI手术可能会改善静态平衡。
确定CI电极插入技术是否对前庭功能有影响。
对36例行标准耳蜗造口术(SC)植入的CI患者和16例行经圆窗(RW)植入的患者在CI手术前后前瞻性地测量静态和动态姿势控制以及冷热前庭功能。测试组合包括改良的平衡感觉交互临床测试(mCTSIB)、节奏性体重转移(RWS)、步行穿过(WA)和串联步行(TW)测试。
大多数CI候选者的冷热测试结果正常,术后仍保持正常。SC组和RW组在CI手术前后的mCTSIB结果无显著差异。在两个亚组中,与术前获得的值相比,mCTSIB综合值在术后6周内有所改善。动态站立平衡在RWS测试中术后无变化。