Baek Dong-Hyun, Lee Eun-Joong, Moon Jin-Hee, Choi Jee Hyun, Pak James Jungho, Lee Sang-Hoon
School of Electrical Engineering and Department of Biomedical Engineering, College of Health Science, Korea University, Jeongneung-dong, Seongbuk-gu, Seoul 136-703, Republic of Korea.
Annu Int Conf IEEE Eng Med Biol Soc. 2009;2009:7022-5. doi: 10.1109/IEMBS.2009.5333320.
In this paper, we have developed 40 channel multiple electrodes mounted on the surface of mouse's skull using polyimide substrate and tested its performance by measuring EEG signals. The recording site of the electrode was electroplated by Pt to enhance both contact impedance and adhesive strength by applying proper current, cleaning surface and removing H(2) gas bubbles. For in vivo test, the electrode was placed on the skull of F1 mouse and EEG signals were measured. We observed the suitability of electrode for measuring EEG signals from multiple areas on the skull. The spectrum of EEG signal to change was observed by urethane administration.
在本文中,我们使用聚酰亚胺基板开发了安装在小鼠颅骨表面的40通道多电极,并通过测量脑电图(EEG)信号来测试其性能。通过施加适当电流、清洁表面和去除氢气气泡,对电极的记录部位进行铂电镀,以提高接触阻抗和粘附强度。对于体内测试,将电极放置在F1小鼠的颅骨上并测量EEG信号。我们观察到该电极对于测量颅骨多个区域的EEG信号具有适用性。通过给予乌拉坦观察到EEG信号变化的频谱。