Oshima Shiori, Sankai Yoshiyuki
Cybernics Laboratory, Graduate School of Systems and Information Engineering, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki, 305-8573, Japan.
Annu Int Conf IEEE Eng Med Biol Soc. 2009;2009:5891-6. doi: 10.1109/IEMBS.2009.5334467.
Blood hematocrit (Hct) is a clinically important index to detect death-dealing symptoms such as anemia and thrombosis, especially for patients with artificial heart, in dialysis and during open-heart surgery. Optical technology has been applied to monitor hematocrit noninvasively, continuously and conveniently, however, it was not well established for actual use. The purpose of this study is to develop an accurate and stable optical hematocrit measurement without any calibrations for device-mounting errors. To this end, we propose a theoretical method. In this method, disturbances are cancelled by using dual detector and optical path changes are calibrated by dual wavelength so that we can measure hematocrit without calibrations of these. Based on the method, a measurement unit that has two LEDs (805/1300[nm]) and four photo detectors was developed. Then, we performed experiments with 38 blood samples from five bovines' blood (Hct: 19-55%). These blood samples were circulated in a mock-up circuit by a blood pump. During the experiment, we measured hematocrit on medical tubing with the developed measurement system. As a result, we could measure hematocrit within 1.7 Hct% mean errors for 38 blood samples without any calibrations. The result indicates that the proposed method is applicable for hematocrit measurement on medical tube in enough small error. We found this proposed method is effective for developments of clinically workable hematocrit measurement/monitoring system.
血液血细胞比容(Hct)是检测贫血和血栓形成等致命症状的重要临床指标,对于人工心脏患者、透析患者以及心脏直视手术患者尤为重要。光学技术已被应用于无创、连续且便捷地监测血细胞比容,然而,其在实际应用中尚未得到完善。本研究的目的是开发一种无需对设备安装误差进行任何校准的准确且稳定的光学血细胞比容测量方法。为此,我们提出一种理论方法。在该方法中,通过使用双探测器消除干扰,并通过双波长校准光路变化,从而无需对这些进行校准即可测量血细胞比容。基于该方法,开发了一种具有两个发光二极管(805/1300[纳米])和四个光电探测器的测量单元。然后,我们使用来自五头牛血液的38个血样(Hct:19 - 55%)进行了实验。这些血样由血泵在模拟回路中循环。在实验过程中,我们使用开发的测量系统在医用管道上测量血细胞比容。结果,我们在未进行任何校准的情况下,对38个血样测量血细胞比容的平均误差在1.7 Hct%以内。结果表明,所提出的方法适用于在医用管道上进行血细胞比容测量,误差足够小。我们发现所提出的方法对于开发临床上可行的血细胞比容测量/监测系统是有效的。