Suppr超能文献

为模块化功能电刺激器实施的安全措施。

Safety measures implemented for modular functioning electrical stimulators.

作者信息

Chen Chiun-Fan, Lai Jin-Shin, Chen Shih-Wei, Lin Yin-Tsong, Kuo Te-Son

机构信息

Department of Electrical Engineering, National Taiwan University, Taipei, Taiwan, R.O.C.

出版信息

Annu Int Conf IEEE Eng Med Biol Soc. 2009;2009:7220-3. doi: 10.1109/IEMBS.2009.5335264.

Abstract

The modular architecture allows for greater flexibility in the building of neural prostheses with a variety of channels but may result in unpredictable accidents under circumstances such as sensor displacements, improper coordination of the connected modules and malfunction of any individual module. A novel fail-safe interface is offered as a solution that puts in place the necessary safety measures when building a module based functional electrical stimulator. By using a single reference line in the interconnecting bus of the modules, various commands would immediately be directed to each module so that proper actions may be taken.

摘要

模块化架构在构建具有多种通道的神经假体时具有更大的灵活性,但在诸如传感器移位、连接模块协调不当以及任何单个模块出现故障等情况下,可能会导致不可预测的事故。本文提出了一种新颖的故障安全接口作为解决方案,该方案在构建基于模块的功能性电刺激器时采取必要的安全措施。通过在模块的互连总线中使用单一参考线,各种命令将立即被导向每个模块,以便采取适当的行动。