Robinson J, Moseley M J, Fielder A R, Bayliss S C
Department of Ophthalmology, University of Birmingham Medical School.
Arch Dis Child. 1991 Jan;66(1 Spec No):59-61. doi: 10.1136/adc.66.1_spec_no.59.
The transmission characteristics of phototherapy eyeshields have been measured under conditions that mimic the clinical situation. Peak transmission (less than 10%) was detected at 700 nm for the poorest patch tested and less than 2% for either of the other patches examined. The optics of measurement systems are considered with reference both to this and previous studies of light transmission through eyepatches. The simplicity and effectiveness with which eyepatches can be secured may be as important as their transmission characteristics.
已在模拟临床情况的条件下测量了光疗眼罩的传输特性。对于测试的最差的眼罩贴片,在700纳米处检测到的峰值传输率(小于10%),而对于所检查的其他任何一个眼罩贴片,峰值传输率均小于2%。测量系统的光学特性参考了此项研究以及之前关于光透过眼罩的研究。眼罩能够固定的简易性和有效性可能与其传输特性同样重要。