Hashida S, Tanaka K, Inoue S, Hayakawa K, Ishikawa E
Department of Biochemistry, Medical College of Miyazaki, Japan.
J Clin Lab Anal. 1991;5(1):38-42. doi: 10.1002/jcla.1860050107.
A specific and sensitive time-resolved fluorometric sandwich immunoassay for human growth hormone (hGH) in serum and urine is described. A monoclonal anti-hGH IgG1 (5802)-coated polystyrene ball was incubated with serum or dialyzed urine and subsequently with europium ion-labeled monoclonal anti-hGH IgG1 (5801). Europium ion bound to the polystyrene ball was measured by time-resolved fluorometry. The detection limit of hGH was 0.3 pg/tube, which was 15-fold higher than that by sandwich enzyme immunoassay using horseradish peroxidase as label. The assay range of serum hGH was 15-15,000 ng/liter using 20 microliters of serum. The assay range of urinary hGH was 2-2,000 ng/liter using 150 microliters of dialyzed urine. The detection limits of hGH in serum and urine by this immunoassay were satisfactory for measuring hGH levels in serum and urine of healthy children and in serum of healthy adults and higher levels but not in urine of healthy adults and in serum and urine of patients with hGH deficiency.
本文描述了一种用于检测血清和尿液中人生长激素(hGH)的特异性和灵敏的时间分辨荧光夹心免疫分析法。将包被有抗hGH单克隆IgG1(5802)的聚苯乙烯球与血清或透析后的尿液孵育,随后再与铕离子标记的抗hGH单克隆IgG1(5801)孵育。通过时间分辨荧光法测量结合在聚苯乙烯球上的铕离子。hGH的检测限为0.3 pg/管,比以辣根过氧化物酶为标记物的夹心酶免疫分析法高15倍。使用20微升血清时,血清hGH的检测范围为15 - 15,000 ng/升。使用150微升透析后的尿液时,尿hGH的检测范围为2 - 2,000 ng/升。该免疫分析法对血清和尿液中hGH的检测限,对于测量健康儿童的血清和尿液以及健康成年人的血清中的hGH水平是令人满意的,对于较高水平的hGH也是如此,但对于健康成年人的尿液以及hGH缺乏患者的血清和尿液则不适用。