Specific shortcomings of endograft design.

作者信息

Yin Tieying, Guidoin Robert, Corriveau Marc M, Nutley Mark, Xu Lu, Marinov Georgi R, Wang Lu, Merhi Yahye, McGregor Robert, Zhang Ze, Douville Yvan, Turgeon Stephane, King Martin, Steinmetz Oren K

机构信息

Department of Surgery, Laval University and Quebec Biomaterials Institute, CHUQ, Quebec, Canada.

出版信息

J Long Term Eff Med Implants. 2008;18(3):181-204. doi: 10.1615/jlongtermeffmedimplants.v18.i3.10.

Abstract

Further to the rapid enlargement of an aneurysm to 5.6 cm in diameter after 3 years of surveillance, a 79-year-old patient was fitted with a Vanguard modular stent graft and monitored on a regular basis for 6 years. Two years later, the aneurysmal sac ruptured. The patient died 1 month after an open surgery. The device was devoid of any encapsulation and the ipsilateral limb was detached from the body. The Nitinol skeleton was mostly maintained, however, some polypropylene sutures were broken. The resulting motion of the sharp-angled Nitinol wires caused abrasion and resulted in a few localized holes that were sufficient to permit blood to percolate through the textile wall. Some polyester yarns in the warp direction were ruptured. The Nitinol wire used in this device was shown to be corrosion resistant but the selection of the polypropylene suture was inappropriate. Because this technology is maturing rapidly, these weaknesses can be avoided in the future generations of endovascular devices. It is recommended that these Nitinol wires be sutured to the fabric and that polyester yarns stronger than 68 decitex in tubes 8 mm in diameter are selected.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验