Gazi University, Department of Chemistry, 06500 Ankara, Turkey.
Talanta. 2010 Jan 15;80(3):1129-32. doi: 10.1016/j.talanta.2009.08.037.
A new solid state fluoride ion selective electrode composed of 70% Ag(2)S, 10% Cu(2)S and 20% CaF(2) has been developed. An analytically useful potential change occurred, from 1x10(-6) to 1x10(-1)M fluoride ion. The slope of the linear portion (1x10(-1)-1x10(-5)M) was about 26+/-2mV/10-fold concentration changes in fluoride. It was found that pH change between 1 and 8 had no effect on the potential of the electrode. There was no interference of most common cations such as K(+,) Na(+), Ca(2+) and Mg(2+) and anions such as Cl(-), NO(3)(-), SO(4)(2-) and PO(4)(3-). The lifetime of the electrode was more than 2 years, when used at least 4-5 times a day, and the response time was about 60s. The measurements were made at constant ionic strength (0.1M NaNO(3)) and at room temperature. This electrode has been used for the determination of fluoride ion in Ankara city tap water and in bottled spring water using standard addition method. The validation of the electrode has been made with a commercial fluoride ion selective electrode (Orion) and high consistency was obtained.
一种由 70%Ag2S、10%Cu2S 和 20%CaF2 组成的新型固态氟离子选择电极已经被开发出来。从 1x10(-6)到 1x10(-1)M 的氟离子浓度,发生了具有分析意义的电位变化。线性部分(1x10(-1)-1x10(-5)M)的斜率约为 26+/-2mV/10 倍氟离子浓度变化。结果发现,电极电位不受 pH 值在 1 到 8 之间变化的影响。大多数常见阳离子,如 K(+), Na(+), Ca(2+) 和 Mg(2+)以及阴离子,如 Cl(-), NO(3)(-), SO(4)(2-) 和 PO(4)(3-),均无干扰。该电极的使用寿命超过 2 年,每天至少使用 4-5 次,响应时间约为 60s。测量在恒定离子强度(0.1M NaNO3)和室温下进行。该电极已用于测定安卡拉市自来水和瓶装矿泉水的氟离子含量,使用标准添加法。通过与商业氟离子选择电极(Orion)进行验证,获得了高度一致的结果。