Division of Endocrinology and Metabolism, Department of Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Head Neck. 2010 Sep;32(9):1161-6. doi: 10.1002/hed.21308.
Serum thyroglobulin (Tg) is essential in the follow-up of patients with differentiated thyroid carcinoma (DTC). However, interchangeability and standardization between Tg assays have not yet been achieved, even with the development of an international Tg standard (Certified Reference Material 457 [CRM-457]).
Serum Tg from 30 DTC patients and serially diluted CRM-457 were measured using 3 different immunoradiometric assays (IRMA-1, IRMA-2, IRMA-3). The intraclass correlation coefficient (ICC) method was used to describe the concordance of each IRMA to CRM-457.
The serum Tg measured by 3 different IRMAs correlated well (r > .85, p < .0001), but clinically relevant discrepancies were found in 13.3% of patients. IRMA-3, which claims to be standardized to CRM-457, showed the best ICC (p(1) = .98) for the CRM-457.
Hospitals caring for patients with DTC should either set their own cutoffs for IRMAs for Tg based on their patient pools, or adopt IRMAs standardized to CRM-457 and calibrate their laboratory using CRM-457.