Yang Jun, Zhen Xin, Lu Wen-ting, Zhou Ling-hong
School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
Nan Fang Yi Ke Da Xue Xue Bao. 2009 Dec;29(12):2379-82.
The FDK algorithm is one of the most widely used algorithms for cone-beam CT reconstruction in circular trajectory due to its simplicity of implementation and computational efficiency. However, the images reconstructed by the FDK algorithm based on real projection data may be blurred without electronic correction and geometric calibration, and are often plagued by deleterious ring artifacts and shading artifacts. We compared the images reconstructed with and without detector correction based on computer experiment of the real biological object. The experimental simulation shows that these algorithms are effective in reducing the ring artifact without compromising the image resolution, and produce satisfactory results.
FDK算法是用于圆形轨迹锥束CT重建的最广泛使用的算法之一,因其实现简单且计算效率高。然而,基于真实投影数据通过FDK算法重建的图像在没有电子校正和几何校准的情况下可能会模糊,并且经常受到有害的环形伪影和阴影伪影的困扰。我们基于真实生物对象的计算机实验,比较了有无探测器校正时重建的图像。实验模拟表明,这些算法在不影响图像分辨率的情况下有效减少环形伪影,并产生了令人满意的结果。