The effect of denture cleansing solutions on the retention of pink Locator attachments: an in vitro study.

机构信息

ASE Prosthodontics, University of Maryland Dental School, Baltimore, MD.

出版信息

J Prosthodont. 2010 Apr;19(3):226-30. doi: 10.1111/j.1532-849X.2009.00550.x. Epub 2009 Dec 21.

Abstract

PURPOSE

To evaluate the changes in retention of pink Locator attachments after exposure to various denture cleansers.

MATERIALS AND METHODS

Six groups (20 pairs each) of pink Locator attachments (3.0 lb. Light Retention replacement patrix attachments) were soaked for the equivalent of 6 months of clinical use in the following solutions: Water (control), Polident Regular, Efferdent, 6.15% sodium hypochlorite (NaOCL, 1:10 dilution), Polident Overnight, and Cool Mint Listerine mouthwash. A universal testing machine set at a crosshead speed of 2 in/min was used to perform one pull. The peak load-to-dislodgement was recorded to reflect changes in the retention of the Locator attachments after soaking. Data were analyzed by one-way ANOVA followed by Tukey's Honestly Significant Difference test. A p< or = 0.05 was considered significant.

RESULTS

Denture cleansing solutions significantly affected the retentive values of pink Locator attachments (F = 344.3, p< or = 0.0001). Cool Mint Listerine mouthwash increased the retentive values of the attachments (51.10 +/- 5.31 N) when compared to the control group (45.25 +/- 3.49 N). There was no significant difference in the retentive values of attachments soaked in Polident Regular or Polident Overnight when compared to the control group. Efferdent caused a small reduction in the retentive values (40.81 +/- 2.56 N) and most importantly, diluted NaOCl caused a large reduction in the retentive values (7.83 +/- 2.50 N) of pink Locator attachments. In addition, Cool Mint Listerine mouthwash caused blue discoloration of the Locator attachments, and NaOCl caused whitening and softening of the pink Locator attachments.

CONCLUSION

Cool Mint Listerine and Efferdent's small effect on the retentive values of the Locators might be clinically unimportant; however, NaOCl caused a large reduction in the retentive values of the attachments. Because of their effect on retentive values and on the color of the Locator attachments, NaOCl and Cool Mint Listerine are not recommended. These results should be interpreted clinically with caution, realizing that different results may be obtained when fatigue stress during function and multiple pulls (in vivo) are combined with the chemical action of denture cleansers.

摘要

目的

评估不同义齿清洁剂对粉色 Locator 附着体固位力保留的影响。

材料和方法

将 6 组(每组 20 对)粉色 Locator 附着体(3.0 lb 轻力固位矩台附着体)浸泡在以下溶液中,相当于临床使用 6 个月:水(对照)、Polident Regular、Efferdent、6.15%次氯酸钠(1:10 稀释)、Polident Overnight 和 Cool Mint Listerine 漱口水。使用十字头速度设置为 2 in/min 的万能试验机进行一次拉伸。记录峰值负载-脱位,以反映浸泡后 Locator 附着体固位力的变化。数据采用单因素方差分析,然后进行 Tukey Honestly Significant Difference 检验。p<0.05 被认为具有统计学意义。

结果

义齿清洁剂显著影响粉色 Locator 附着体的固位值(F = 344.3,p<0.0001)。与对照组相比,Cool Mint Listerine 漱口水增加了附着体的固位值(51.10 +/- 5.31 N)。与对照组相比,Polident Regular 或 Polident Overnight 浸泡的附着体固位值无显著差异。Efferdent 导致附着体固位值略有下降(40.81 +/- 2.56 N),最重要的是,稀释的次氯酸钠导致粉色 Locator 附着体固位值大幅下降(7.83 +/- 2.50 N)。此外,Cool Mint Listerine 漱口水导致 Locator 附着体出现蓝色变色,次氯酸钠导致粉色 Locator 附着体变白和变软。

结论

Cool Mint Listerine 和 Efferdent 对 Locator 固位力的微小影响可能在临床上无足轻重;然而,次氯酸钠大大降低了附着体的固位力。由于对固位力和 Locator 附着体颜色的影响,不建议使用次氯酸钠和 Cool Mint Listerine。这些结果应谨慎地从临床角度进行解释,同时意识到在功能疲劳应力和多次拉伸(体内)与义齿清洁剂的化学作用相结合时,可能会得到不同的结果。