Department of Chemistry and Biochemistry, University of Texas, 700 Planetarium Place, Arlington, Texas 76019-0065, USA.
Anal Chem. 2010 Feb 1;82(3):951-8. doi: 10.1021/ac9021902.
We describe a flow-through ionic charge detector in the form of a three-compartmented system. A central water channel is separated from two outer channels bearing water (or a dilute electrolyte) by a cation-exchange membrane (CEM) and an anion-exchange membrane (AEM). Independent fluid input/output ports address all channels. One platinum electrode is put in each outer channel. When the AEM-side electrode is positive with respect to the CEM-side electrode and voltage (approximately 1-10 V) is applied, the observed background current is from the transport of H(+)/OH(-) through the CEM/AEM to the negative/positive electrodes, respectively. The H(+) and OH(-) are generated by the ionization of water, in part aided by the electric field. If an electrolyte (X(+)Y(-)) is injected in to the central channel, X(+) and Y(-) migrate through the CEM and AEM to the negative and positive electrodes, respectively, and generate a current pulse. The integrated area of the current signal (coulombs) elicited by this electrolyte injection is dependent on a number of variables, the most important being the central channel residence time and the applied voltage (V(app)); these govern the transport of the injected electrolyte to/through the membranes. Other parameters include electrode placement, fluid composition, and outer channel flow rates. For strong electrolytes, depending on the operating conditions, the current peak area (hereinafter called the measured charge signal, Q(m)) can both be less or more than the charge represented by the electrolyte injected (Q(i)). Q(m) is less than Q(i) if transport to/through the membranes is subquantitative. Q(m) can be greater than Q(i) at higher V(app). At constant V(app) more voltage is dropped across the membranes as the central channel becomes more conductive and water dissociation at the membrane surface is enhanced. Effectively, the membranes experience a greater applied voltage as the central channel becomes more conductive. The resulting additional current accompanying analyte introduction to the detector can substantially augment Q(m). Thus, the device is not an absolute coulometer although V(app) can be deliberately chosen to have Q(m) = Q(i) over at least a 10-fold concentration range. Importantly, equivalent amounts of diverse strong electrolytes (with substantially different conductivities) injected into the central channel produce the same charge signals. In ion chromatography, this results in identical calibration curves for all strong acid anions, obviating individual calibrations. Whereas with a conductivity detector (CD) only the ionized portion of a weak electrolyte responds, in the present charge detector (ChD), ions are actually removed, leading to further ionization and the detection of a proportionately greater analyte amount.
我们描述了一种以三隔室系统形式存在的流通式离子电荷检测器。中央水道与两个外部通道隔开,外部通道分别承载水(或稀电解质),并由阳离子交换膜(CEM)和阴离子交换膜(AEM)隔开。独立的流体输入/输出端口可用于所有通道。每个外部通道中都有一个铂电极。当 AEM 侧电极相对于 CEM 侧电极为正且施加电压(约 1-10V)时,观察到的背景电流来自于 H(+) / OH(-) 通过 CEM/AEM 分别向负/正电极的传输。H(+)和 OH(-)是由水的部分电离产生的,部分得益于电场的作用。如果将电解质(X(+)Y(-))注入中央通道,X(+)和 Y(-)将分别通过 CEM 和 AEM 迁移到负电极和正电极,并产生电流脉冲。由这种电解质注入引起的电流信号(库仑)的积分面积取决于许多变量,最重要的是中央通道停留时间和施加电压(V(app));这些变量控制着注入的电解质向/通过膜的传输。其他参数包括电极位置、流体组成和外通道流速。对于强电解质,根据操作条件的不同,电流峰面积(以下称为测量电荷信号,Q(m))可能小于或大于注入的电解质(Q(i))所代表的电荷。如果向/通过膜的传输是亚定量的,则 Q(m)小于 Q(i)。在较高的 V(app)下,Q(m)可以大于 Q(i)。随着中央通道的导电性增加,在膜表面的水离解增强,更多的电压降落在膜上。实际上,随着中央通道的导电性增加,膜会受到更大的施加电压。伴随着分析物引入检测器的额外电流会显著增加 Q(m)。因此,尽管可以故意选择 V(app)以使 Q(m)=Q(i)至少在 10 倍浓度范围内,但该设备不是绝对库仑计。重要的是,注入中央通道的等量不同强电解质(具有显著不同的电导率)会产生相同的电荷信号。在离子色谱中,这会导致所有强酸阴离子的校准曲线相同,无需单独校准。而在电导检测器(CD)中,只有弱电解质的电离部分有响应,而在本电荷检测器(ChD)中,实际上是去除了离子,导致进一步电离,并检测到比例更大的分析物量。