Dias Mauro S, Tongu Margareth L O, Takeda Mauro N, Koskinas Marina F
Instituto de Pesquisas Energéticas e Nucleares, IPEN-CNEN/SP, Centro do Reator de Pesquisas-CRPq, São Paulo, SP, Brazil.
Appl Radiat Isot. 2010 Jul-Aug;68(7-8):1362-6. doi: 10.1016/j.apradiso.2009.12.022. Epub 2009 Dec 21.
The present work describes the methodology for predicting the behavior of extrapolation curves obtained in radionuclide standardization by 4pibeta-gamma coincidence measurements, applied to (22)Na, developed at the Laboratório de Metrologia Nuclear of IPEN-CNEN/SP (LMN-Nuclear Metrology Laboratory). The LMN system consists of a proportional counter (PC) in 4pi geometry coupled to a single or a pair of NaI(Tl) scintillation crystals. Two standardization techniques were used: the Sum-Peak and the Nuclear-Peak methods. The theoretical response functions of each detector have been calculated using the MCNPX Monte Carlo code. The code ESQUEMA, developed at LMN, has been used for calculating the extrapolation curve in the 4pibeta-gamma coincidence experiment. Modifications were performed in order to include response tables for positrons and coincidences with annihilation photons. From the calibration results it was possible to extract both the activity value and the positron emission probability per decay. The latter was compared with results from the literature.