McGovern John-Paul, Shih Wei-Heng, Rest Richard F, Purohit Mitali, Mattiucci Mark, Pourrezaei Kambiz, Onaral Banu, Shih Wan Y
Department of Materials Science and Engineering, Drexel University, Philadelphia, Pennsylvania 19104, USA.
Rev Sci Instrum. 2009 Dec;80(12):125104. doi: 10.1063/1.3264082.
An array of three identical piezoelectric microcantilever sensors (PEMSs) consisting of a lead zirconate titanate layer bonded to a glass layer was fabricated and examined for simultaneous, in situ, real-time, all-electrical detection of Bacillus anthracis (BA) spores in an aqueous suspension using the first longitudinal extension mode of resonance. With anti-BA antibody immobilized on the sensor surfaces all three PEMS exhibited identical BA detection resonance frequency shifts at all tested concentrations, 10-10(7) spores/ml with a standard deviation of less than 10%. The detection concentration limit of 10 spores/ml was about two orders of magnitude lower than would be permitted by flexural peaks. In blinded-sample testing, the array PEMS detected BA in three samples containing BA: (1) 3.3x10(3) spores/ml, (2) a mixture of 3.3x10(3) spores/ml and 3.3x10(5) S. aureus (SA) and P. aeruginosa (PA) per ml, and (3) a mixture of 3.3x10(3) spores/ml with 3.3x10(6) SA+PA/ml. There was no response to a sample containing only 3.3x10(6) SA+PA/ml. These results illustrate the sensitivity, specificity, reusability, and reliability of array PEMS for in situ, real-time detection of BA spores.
制备了一组由三个相同的压电微悬臂梁传感器(PEMS)组成的阵列,该传感器由键合到玻璃层上的锆钛酸铅层构成,并使用共振的第一纵向扩展模式对水悬浮液中的炭疽芽孢杆菌(BA)孢子进行同步、原位、实时、全电检测。在传感器表面固定抗BA抗体后,所有三个PEMS在所有测试浓度(10 - 10⁷个孢子/毫升)下均表现出相同的BA检测共振频率偏移,标准偏差小于10%。10个孢子/毫升的检测浓度极限比弯曲峰允许的浓度极限低约两个数量级。在盲样测试中,阵列PEMS在三个含有BA的样品中检测到了BA:(1)3.3×10³个孢子/毫升,(2)每毫升含有3.3×10³个孢子以及3.3×10⁵个金黄色葡萄球菌(SA)和铜绿假单胞菌(PA)的混合物,(3)每毫升含有3.3×10³个孢子以及3.3×10⁶个SA + PA的混合物。对仅含有3.3×10⁶个SA + PA/毫升的样品没有响应。这些结果说明了阵列PEMS用于原位、实时检测BA孢子的灵敏度、特异性、可重复使用性和可靠性。