Lipsett F R, Bechthold G, Blair F D, Cairns F V, O'Hara D H
Appl Opt. 1970 Jun 1;9(6):1312-8. doi: 10.1364/AO.9.001312.
An apparatus is designed for use in obtaining fluorescence spectra, excitation spectra, and delayed fluorescence spectra, usually at low temperatures. A photometrically regulated 6500-W xenon lamp is used as excitation source, a double monochromator or varioilluminator as excitation monochromator; a variable temperature liquid helium cryostat as specimen holder; a double grating monochromator or varioilluminator as detection monochromator; and a photomultiplier as detector. A quartz-iodine lamp calibrated against a standard tungsten-quartz lamp is used for intensity calibrations. Fluorescence spectra are recorded by a digital system with a shaft position encoder for digitizing wavelength and an analog-to-digital converter for intensity. Data are recorded on magnetic tape which can be processed directly by an IBM system 360 computer. The selection of components of adequate resolution, speed, and capacity is discussed, and the computer programs are described. Installation of the recording system 1ms facilitated greater wavelength accuracy and long calculations on the experimental data. Examples are given.
设计了一种通常在低温下用于获取荧光光谱、激发光谱和延迟荧光光谱的仪器。使用一台光度调节的6500瓦氙灯作为激发源,一台双单色仪或可变照明器作为激发单色仪;一个可变温度的液氦低温恒温器作为样品架;一台双光栅单色仪或可变照明器作为检测单色仪;以及一个光电倍增管作为检测器。使用一台根据标准钨石英灯校准的石英碘灯进行强度校准。荧光光谱由一个数字系统记录,该系统带有一个用于波长数字化的轴位置编码器和一个用于强度的模数转换器。数据记录在磁带上,可由IBM系统360计算机直接处理。讨论了选择具有足够分辨率、速度和容量的组件,并描述了计算机程序。记录系统的安装有助于提高波长精度并对实验数据进行长时间计算。给出了示例。