Biocompatibility of extracorporeal circulation. In vitro comparison of heparin-coated and uncoated oxygenator circuits.

作者信息

Videm V, Mollnes T E, Garred P, Svennevig J L

机构信息

Institute for Experimental Medical Research, Ullevaal Hospital, Oslo, Norway.

出版信息

J Thorac Cardiovasc Surg. 1991 Apr;101(4):654-60.

PMID:2008104
Abstract

Oxygenator/tubing sets coated with endpoint-attached heparin were compared to uncoated sets in a dynamic model of extracorporeal circulation. Biocompatibility was assessed by evaluations of complement activation and platelet loss. The median concentration of C3 activation products increased gradually from 12 AU/ml at baseline to 65 AU/ml after 120 minutes in the uncoated sets and from 12 AU/ml to 19 AU/ml after 120 minutes in the coated sets (p less than 0.002). The median concentration of the terminal complement complex in the uncoated sets increased gradually from 3.1 AU/ml at baseline to 18.2 AU/ml after 120 minutes. In the coated sets the terminal complement complex reached a peak of 12.0 AU/ml at 15 minutes and returned to baseline values at 60 minutes. Median platelet loss at 120 minutes was 166 x 10(9) in the uncoated and 21 x 10(9) in the coated sets (p less than 0.01). Heparin coating thus improved biocompatibility by reducing both complement activation and platelet loss.