Cancer Prevention and Control Division, Moffitt Cancer Center, 12902 Magnolia Drive, Tampa, Florida 33612, USA.
Med Phys. 2009 Dec;36(12):5380-90. doi: 10.1118/1.3250842.
Breast density is a significant breast cancer risk factor. Although various methods are used to estimate breast density, there is no standard measurement for this important factor. The authors are developing a breast density standardization method for use in full field digital mammography (FFDM). The approach calibrates for interpatient acquisition technique differences. The calibration produces a normalized breast density pixel value scale. The method relies on first generating a baseline (BL) calibration dataset, which required extensive phantom imaging. Standardizing prospective mammograms with calibration data generated in the past could introduce unanticipated error in the standardized output if the calibration dataset is no longer valid.
Sample points from the BL calibration dataset were imaged approximately biweekly over an extended timeframe. These serial samples were used to evaluate the BL dataset reproducibility and quantify the serial calibration accuracy. The cumulative sum (Cusum) quality control method was used to evaluate the serial sampling.
There is considerable drift in the serial sample points from the BL calibration dataset that is x-ray beam dependent. Systematic deviation from the BL dataset caused significant calibration errors. This system drift was not captured with routine system quality control measures. Cusum analysis indicated that the drift is a sign of system wear and eventual x-ray tube failure.
The BL calibration dataset must be monitored and periodically updated, when necessary, to account for sustained system variations to maintain the calibration accuracy.
乳房密度是乳腺癌的一个重要危险因素。虽然有多种方法用于估计乳房密度,但对于这个重要因素还没有标准的测量方法。作者正在为全视野数字化乳腺摄影(FFDM)开发一种乳房密度标准化方法。该方法针对患者间的采集技术差异进行校准。校准会生成归一化的乳房密度像素值范围。该方法依赖于首先生成基线(BL)校准数据集,这需要进行广泛的体模成像。如果校准数据集不再有效,使用过去生成的校准数据对前瞻性乳房 X 光片进行标准化可能会导致标准化输出中出现意外错误。
BL 校准数据集中的采样点在较长时间内大约每两周进行一次成像。这些连续样本用于评估 BL 数据集的可重复性,并量化连续校准的准确性。累积和(Cusum)质量控制方法用于评估连续采样。
BL 校准数据集中的连续采样点存在明显的漂移,这种漂移与 X 射线束有关。与 BL 数据集的系统偏差导致了显著的校准误差。这种系统漂移无法通过常规系统质量控制措施捕捉到。Cusum 分析表明,漂移是系统磨损和最终 X 射线管故障的迹象。
必须对 BL 校准数据集进行监测,并在必要时进行定期更新,以应对持续的系统变化,从而保持校准精度。