Moss G E, Miller L R, Forward R L
Appl Opt. 1971 Nov 1;10(11):2495-8. doi: 10.1364/AO.10.002495.
We have constructed and tested a laser interferometer transducer for a long, wideband, laser-linked gravitational radiation antenna. Photon-noise-limited performance was achieved using 80 microW from a single mode Spectra-Physics 119 laser in a modified Michelson interferometer on a vibration isolation table in a quiet room. A piezoelectric driver on one of the interferometer mirrors was used to generate subangstrom (3 x 10(-14)-m) vibrations of known amplitude. The measured displacement sensitivity of the system in the kilohertz region was 1.3 x 10(-14)m/Hz(1/2), which compares well with the calculated photon noise limit of 1.06 x 10(-14) m/Hz(1/2). This is the smallest vibrational displacement measured directly with a laser to date.