Jahrome A Kh, Stella Peter R, Leijdekkers Vanessa J, Guyomi Siyrous Hoseyni, Moll Frans L
Department of Vascular Surgery, University Medical Centre Utrecht, Utrecht, the Netherlands.
Vascular. 2010 Jan-Feb;18(1):59-61. doi: 10.2310/6670.2009.00026.
A 41-year-old woman was treated with a Figulla (Occlutec, Helsingborg, Sweden) atrial septum occluder device with no intraprocedural complications. Five months later, dislocation of the device in the abdominal aorta was detected. The occluder device was located at the level of the celiac axis, nearly obstructing the entire aorta. Owing to total incorporation of the device, endoluminal retrieval was not possible. Through a medial rotation approach, the device was safely removed. This is a rare complication after endoluminal closure of an atrial septum defect. The retrieval possibilities are discussed.
一名41岁女性接受了菲古拉(瑞典赫尔辛堡奥克鲁泰克公司生产)房间隔封堵器治疗,术中无并发症。五个月后,发现封堵器移位至腹主动脉。封堵器位于腹腔干水平,几乎阻塞了整个主动脉。由于封堵器完全嵌入,无法通过腔内取出。通过内侧旋转方法,该装置被安全取出。这是房间隔缺损腔内封堵术后罕见的并发症。文中讨论了取出的可能性。