Optom Vis Sci. 2010 Mar;87(3):195-9. doi: 10.1097/OPX.0b013e3181d1d940.
To study the usefulness of the Waveform Score from the ocular response analyzer (ORA).
Both eyes of sixty-four normal Chinese adults were measured by the ORA. An experienced practitioner who was masked to the score evaluated the waveforms. Four measurements were obtained from each eye, and the average was considered as the gold standard. Agreement was compared for the mean of different multiple measurements to the gold standard.
There was no significant difference between the gold standard and the mean of different multiple measurements for all ORA parameters. The Waveform Score of 512 signals (four measurements from each eye of 64 subjects) ranged from 1.58 to 9.06. When the best signal value of four measured signals from each eye was considered, the lowest score was 3.60 from 128 eyes. The lower 10th percentile from all signals had Waveform Scores <3.48.
If the Waveform Score provides information on the reliability of the signals, a score <3.50 may indicate an unreliable signal, and the signal should be discarded. We also recommend taking three measurements with all the signals having a Waveform Score of 3.50 or above to increase the precision.
研究眼反应分析仪(ORA)的波形评分的实用性。
对64名正常中国成年人的双眼进行ORA测量。一名对评分不知情的经验丰富的从业者评估波形。每只眼睛进行四次测量,其平均值被视为金标准。将不同多次测量的平均值与金标准进行一致性比较。
所有ORA参数的金标准与不同多次测量的平均值之间无显著差异。512个信号(64名受试者每只眼睛的四次测量)的波形评分范围为1.58至9.06。当考虑每只眼睛四次测量信号中的最佳信号值时,128只眼睛的最低评分为3.60。所有信号中较低的第10百分位数的波形评分<3.48。
如果波形评分提供了信号可靠性的信息,评分<3.50可能表明信号不可靠,该信号应被舍弃。我们还建议对所有波形评分在3.50及以上的信号进行三次测量,以提高精度。